Comparisons of disposable and conventional silver working electrode for the determination of iodide using high-performance anion-exchange chromatography with pulsed amperometric detection
文献类型:期刊论文
作者 | Liang, LN; Cai, YQ; Mou, SF; Cheng, J |
刊名 | JOURNAL OF CHROMATOGRAPHY A
![]() |
出版日期 | 2005 |
卷号 | 1085期号:1页码:37-41 |
关键词 | disposable silver working electrode conventional silver working electrode iodide HPAEC PAD |
英文摘要 | The paper compared the performance of two kinds of silver working electrode in electrochemical detector-the disposable and conventional electrode for the determination of iodide using high-performance anion-exchange chromatography (HPAEC) hyphenated with pulsed amperometric detection (PAD). The comparisons were carried out on the time of equilibration, long and short-term reproducibility, limits of detection and linearity of calibration. Results showed that disposable working electrode manifested equal or better results than conventional working electrode and could be used for iodide analysis. Besides, the disposable electrode could work for consecutive 2660 min (about 44 h, 10 min needed for each run) with no degradation. Due to its "disposable" property, disposable working electrode could be discarded if the detection sensitivity decreased to 80% so the time for polishing and reconditioning was spared and good reproducible results could be obtained. At last, the disposable electrode was applied for the determination of iodide in soil and sea water samples with the spiked recovery ranging from 96-104% and the detection limit of 0.5 mu g/L (10 mu L injection, three times of the baseline noise). (c) 2004 Elsevier B.V. All rights reserved. |
学科主题 | Biochemistry & Molecular Biology; Chemistry |
研究领域[WOS] | Biochemistry & Molecular Biology ; Chemistry |
公开日期 | 2015-12-24 |
源URL | [http://ir.rcees.ac.cn/handle/311016/23156] ![]() |
专题 | 生态环境研究中心_环境化学与生态毒理学国家重点实验室 |
推荐引用方式 GB/T 7714 | Liang, LN,Cai, YQ,Mou, SF,et al. Comparisons of disposable and conventional silver working electrode for the determination of iodide using high-performance anion-exchange chromatography with pulsed amperometric detection[J]. JOURNAL OF CHROMATOGRAPHY A,2005,1085(1):37-41. |
APA | Liang, LN,Cai, YQ,Mou, SF,&Cheng, J.(2005).Comparisons of disposable and conventional silver working electrode for the determination of iodide using high-performance anion-exchange chromatography with pulsed amperometric detection.JOURNAL OF CHROMATOGRAPHY A,1085(1),37-41. |
MLA | Liang, LN,et al."Comparisons of disposable and conventional silver working electrode for the determination of iodide using high-performance anion-exchange chromatography with pulsed amperometric detection".JOURNAL OF CHROMATOGRAPHY A 1085.1(2005):37-41. |