A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses
文献类型:期刊论文
作者 | Longo, M; Rigon, L; Lopez, FCM; Chen, RC; Dreossi, D; Zanconati, F; Longo, R |
刊名 | PHYSICS IN MEDICINE AND BIOLOGY
![]() |
出版日期 | 2015 |
卷号 | 60期号:3页码:N21—N34 |
关键词 | IMAGING TECHNIQUE RAY TOMOGRAPHY |
英文摘要 | This work presents the first study of x-ray phase contrast imaging based on a simple implementation of the edge illumination method (EIXPCi) in the field of mammography with synchrotron radiation. A simplified EIXPCi set-up was utilized to study a possible application in mammography at clinical doses. Moreover, through a novel algorithm capable of separating and quantifying absorption and phase perturbations of images acquired in EIXPCi modality, it is possible to extract quantitative information on breast images, allowing an accurate tissue identification. The study was carried out at the SYRMEP beamline of Elettra synchrotron radiation facility (Trieste, Italy), where a mastectomy specimen was investigated with the EIXPCi technique. The sample was exposed at three different energies suitable for mammography with synchrotron radiation in order to test the validity of the novel algorithm in extracting values of linear attenuation coefficients integrated over the sample thickness. It is demonstrated that the quantitative data are in good agreement with the theoretical values of linear attenuation coefficients calculated on the hypothesis of the breast with a given composition. The results are promising and encourage the current efforts to apply the method in mammography with synchrotron radiation. |
收录类别 | SCI |
语种 | 英语 |
公开日期 | 2015-12-24 |
源URL | [http://ir.sinap.ac.cn/handle/331007/24386] ![]() |
专题 | 上海应用物理研究所_中科院上海应用物理研究所2011-2017年 |
推荐引用方式 GB/T 7714 | Longo, M,Rigon, L,Lopez, FCM,et al. A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses[J]. PHYSICS IN MEDICINE AND BIOLOGY,2015,60(3):N21—N34. |
APA | Longo, M.,Rigon, L.,Lopez, FCM.,Chen, RC.,Dreossi, D.,...&Longo, R.(2015).A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses.PHYSICS IN MEDICINE AND BIOLOGY,60(3),N21—N34. |
MLA | Longo, M,et al."A simplified edge illumination set-up for quantitative phase contrast mammography with synchrotron radiation at clinical doses".PHYSICS IN MEDICINE AND BIOLOGY 60.3(2015):N21—N34. |
入库方式: OAI收割
来源:上海应用物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。