中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
A One-Step Immunoassay for Carbohydrate Antigen 19-9 By Biosensor Based on Imaging Ellipsometry

文献类型:期刊论文

作者Huang CH; Chen YY(陈艳艳); Jin G(靳刚)
刊名ANNALS OF BIOMEDICAL ENGINEERING
出版日期2011
卷号39期号:1页码:185-192
通讯作者邮箱huang1ng@tom.com ; yychen2006@sinano.ac.cn ; gajin@imech.ac.cn
关键词Biosensor Imaging ellipsometry Carbohydrate antigen 19-9 Immunoassay
ISSN号0090-6964
产权排序[Huang, Chenghong; Chen, Yanyan; Jin, Gang] Chinese Acad Sci, Suzhou Inst Nanotech & Nanobion, Suzhou 215123, Peoples R China; [Huang, Chenghong; Jin, Gang] Chinese Acad Sci, Inst Biophys, Beijing 100101, Peoples R China; [Jin, Gang] CAS, Inst Mech, Key Lab Micrograv, Beijing 100190, Peoples R China; [Huang, Chenghong] Chinese Acad Sci, Grad Sch, Beijing, Peoples R China
通讯作者Chen, YY (reprint author), Chinese Acad Sci, Suzhou Inst Nanotech & Nanobion, Suzhou 215123, Peoples R China.
中文摘要Cancer markers can offer much more information for developing or developed cancers. Simple and short-time assay of cancer biomarker is essential for clinic diagnosis and management. In this study, a one-step immunoassay for carbohydrate antigen 19-9 (CA19-9) by biosensor based on imaging ellipsometry (BIE) is described and its analysis performances are evaluated. Anti-CA19-9 monoclonal antibody as ligand covalently immobilized on carboxyl-modified silicon substrate and its immobilization concentration concurrently screened. Then, bovine serum albumin (BSA) is used to block for acquisition of sensing layer. The detection linear range is from 15.0 to 120.0 U/mL with a minimum limit of 10.0 U/mL (S/N = 3). The intra-slide and inter-slide reproducibility, expressed as coefficient variation (CV%), are 9.9, 5.9, 5.4% and 9.6, 7.6, 5.3% for 31.9, 57.4, and 99.2 U/mL of CA19-9, respectively. Accuracy determined by spiked-recovery is among 95.1-106.7%. Specificity expressed as cross-reaction rate are all below 5.8% evaluated by three analogous biomarkers alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), and carbohydrate antigen 242 (CA242). Application of BIE to 25 clinic samples demonstrated that the results are in high correlate with electrochemiluminescent immunoassay (ECLIA, R-2 = 0.997). The proposed immunoassay has the potential for clinic application.
分类号二类/Q2
类目[WOS]Engineering, Biomedical
研究领域[WOS]Engineering
关键词[WOS]OPTICAL PROTEIN-CHIP ; SPECTROSCOPIC ELLIPSOMETRY ; COLORECTAL-CANCER ; IMMOBILIZATION ; IMMUNOSENSOR ; MICROARRAYS ; SURFACE ; VISUALIZATION ; LAYERS ; ASSAY
收录类别SCI ; EI
语种英语
WOS记录号WOS:000287213100017
源URL[http://dspace.imech.ac.cn/handle/311007/58591]  
专题力学研究所_国家微重力实验室
推荐引用方式
GB/T 7714
Huang CH,Chen YY,Jin G. A One-Step Immunoassay for Carbohydrate Antigen 19-9 By Biosensor Based on Imaging Ellipsometry[J]. ANNALS OF BIOMEDICAL ENGINEERING,2011,39(1):185-192.
APA Huang CH,陈艳艳,&靳刚.(2011).A One-Step Immunoassay for Carbohydrate Antigen 19-9 By Biosensor Based on Imaging Ellipsometry.ANNALS OF BIOMEDICAL ENGINEERING,39(1),185-192.
MLA Huang CH,et al."A One-Step Immunoassay for Carbohydrate Antigen 19-9 By Biosensor Based on Imaging Ellipsometry".ANNALS OF BIOMEDICAL ENGINEERING 39.1(2011):185-192.

入库方式: OAI收割

来源:力学研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。