Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device
文献类型:期刊论文
| 作者 | Gao L(高龙); Yu BX(俞伯祥) ; Ding YY(丁雅韵) ; Zhou L(周莉) ; Wen LJ(温良剑) ; Xie YG(谢宇广) ; Wang ZG(王志刚) ; Cai X(蔡啸) ; Sun XL(孙希磊) ; Fang J(方建)
|
| 刊名 | Chinese Physics C
![]() |
| 出版日期 | 2013 |
| 期号 | 7页码:53-57 |
| 关键词 | attenuation length liquid scintillator (LS) Beer-Lambert law LabVIEW |
| 通讯作者 | 高龙 |
| 英文摘要 | An attenuation length measurement device was constructed using an oscilloscope and LabVIEW for signal acquisition and processing. The performance of the device has been tested in a variety of ways. The test results show that the set-up has a good stability and high precision (sigma/mean reached 0.4 ... |
| 公开日期 | 2016-02-26 |
| 源URL | [http://ir.ihep.ac.cn/handle/311005/219128] ![]() |
| 专题 | 高能物理研究所_实验物理中心 高能物理研究所_粒子天体物理中心 高能物理研究所_多学科研究中心 |
| 推荐引用方式 GB/T 7714 | Gao L,Yu BX,Ding YY,et al. Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device[J]. Chinese Physics C,2013(7):53-57. |
| APA | 高龙.,俞伯祥.,丁雅韵.,周莉.,温良剑.,...&Wen Liang-Jian.(2013).Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device.Chinese Physics C(7),53-57. |
| MLA | 高龙,et al."Attenuation length measurements of a liquid scintillator with LabVIEW and reliability evaluation of the device".Chinese Physics C .7(2013):53-57. |
入库方式: OAI收割
来源:高能物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。