中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
The study and design of the large area neutron monitor with RPC-Gd

文献类型:期刊论文

作者Qian S(钱森); Ning Z(宁哲); Zhang JW(张家文); Qian, S; Ning, Z; Zhang, J; Zhang, Q; Han, R
刊名JOURNAL OF INSTRUMENTATION
出版日期2015
卷号10页码:C02014
关键词Gaseous detectors Neutron detectors (cold thermal fast neutrons)
通讯作者钱森
英文摘要Thermal neutron absorber gadolinium (Gd) with a large reaction cross section is applied to resistive plate chamber detectors, so a gaseous detector RPC-Gd is developed to discriminate thermal neutrons. Through analyzing the operation principle of RPC-Gd detectors and testing n/gamma spectrum, a method by supplying different operation high voltage of RPC module is used for discriminating thermal neutrons and gamma. In the end, based on the production technology and operation principle of RPC, two prototypes of single/double RPC modules are designed for producing large area and low cost detector to monitor thermal neutron.
学科主题Instruments & Instrumentation
类目[WOS]Instruments & Instrumentation
收录类别SCI
WOS记录号WOS:000351664800014
公开日期2016-05-03
源URL[http://ir.ihep.ac.cn/handle/311005/228268]  
专题中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Qian S,Ning Z,Zhang JW,et al. The study and design of the large area neutron monitor with RPC-Gd[J]. JOURNAL OF INSTRUMENTATION,2015,10:C02014.
APA 钱森.,宁哲.,张家文.,Qian, S.,Ning, Z.,...&Han, R.(2015).The study and design of the large area neutron monitor with RPC-Gd.JOURNAL OF INSTRUMENTATION,10,C02014.
MLA 钱森,et al."The study and design of the large area neutron monitor with RPC-Gd".JOURNAL OF INSTRUMENTATION 10(2015):C02014.

入库方式: OAI收割

来源:高能物理研究所

浏览0
下载0