An Improved QRS Detection Approach based on Two-Lead Bi-Referring
文献类型:会议论文
作者 | Jun Dong(董军)![]() |
出版日期 | 2012 |
会议名称 | The IASTED 2012 African Conferences |
会议日期 | 2012-9-3 |
会议地点 | University of Botswana |
关键词 | QRS Detection Multi-lead ECG R Amplitude Noise Amplitude Signal to Noise Ratio |
英文摘要 | For improving QRS detection accuracy of single lead algorithms in multi-lead electrocardiogram (ECG), based on two-lead bi-referring, a new approach is introduced. As we know, there is much bursting noise in ECG and when it occurs in the detecting lead, it may decrease the accuracy of single lead QRS detecting algorithms. Meanwhile, in clinic, most of ECG is multi-lead, for example the famous MIT-BIH ECG database (MIT-BIH) is two leads system. So, in multi-lead system, it needs to be investigated how to improve existing QRS detection algorithms which are designed only with one lead ECG. In this paper, the final QRS positions are selected from the detection results of the cleanest lead which has least bursting noise and most regular r-peaks. It had been tested in MIT-BIH and Chinese cardiovascular disease database (CCDD) with four single lead algorithms and the results showed that it could improve their accuracy. |
语种 | 英语 |
源URL | [http://58.210.77.100/handle/332007/1077] ![]() |
专题 | 苏州纳米技术与纳米仿生研究所_学科交叉综合研究部_董军团队 |
推荐引用方式 GB/T 7714 | Jun Dong. An Improved QRS Detection Approach based on Two-Lead Bi-Referring[C]. 见:The IASTED 2012 African Conferences. University of Botswana. 2012-9-3. |