Uniformity calibration for ICT image

文献类型:期刊论文

作者Ceng G(曾钢); Liu L(刘力); Que JM(阙介民); Zhang YP(张英平); Yin Y(印胤); Wang YF(王燕芳); Yu ZQ(郁忠强); Yan YL(阎永廉); Zeng, G; Liu, L
刊名HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION
出版日期2004
卷号28期号:10页码:1131-1134
关键词image uniformity beam hardening detector array calibration
通讯作者Chinese Acad Sci, Inst High Energy Phys, Beijing 100039, Peoples R China
英文摘要The uniformity of ICT image is impaired by beam hardening and the inconsistency of detector units responses. This paper analyzes the beam hardening and the nonlinearity of the detector's output. The correction factors are determined experimentally by the detector's responses with different absorption length. The artifacts in the CT image of a symmetrical aluminium cylinder have been eliminated after calibration.