Calibration of stray light based on point source transmittance measurement system

文献类型:期刊论文

作者Xu, Liang1,2; Gao, Li-Min1; Zhao, Jian-Ke1; Liu, Feng1; Zhou, Yan1; Li, Zhao-Hui1,2; Yang, Fei3; Zhao, Qing1,2
刊名guangxue jingmi gongcheng/optics and precision engineering
出版日期2016-07-01
卷号24期号:7页码:1607-1614
关键词Ability testing Distribution functions Equipment Stray light
ISSN号1004924x
其他题名基于点源透过率测试系统的杂散光标定
产权排序1
通讯作者xu, liang (xuliang757@163.com)
英文摘要to improve the stray light testing ability and calibration accuracy of the stray light equipment used in test of the point source transmittance (pst), a calibration lens was proposed to calibrate the testing range and testing accuracy of the equipment in a large off;axial angle. by using a simple physical model, the calibration lens was designed in a laboratory. the physical parameters of the calibration lens were measured, then these parameters were taken into the tracepro to calculate the psts in different off;axial angles. finally, the calculated result and the tested result of psts were compared, and the test accuracy of the equipment were obtained. the experimental results indicate that the difference between calculate results and testing results of the calibration lens is better than lg/0.5, meeting the need of testing accuracy calibration of the stray light equipment in testing psts and providing a reliable reference for the absolute measurement of psts. this technology solves the problem in testing accuracy calibration of pst measuring systems. © 2016, science press. all right reserved.
收录类别EI
语种中文
源URL[http://ir.opt.ac.cn/handle/181661/28256]  
专题西安光学精密机械研究所_检测技术研究中心
作者单位1.Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an; 710119, China
2.University of Chinese Academy of Sciences, Beijing; 100049, China
3.AVIC Xi'an Aircraft Industry (Group) Company LTD., Xi'an; 710089, China
推荐引用方式
GB/T 7714
Xu, Liang,Gao, Li-Min,Zhao, Jian-Ke,et al. Calibration of stray light based on point source transmittance measurement system[J]. guangxue jingmi gongcheng/optics and precision engineering,2016,24(7):1607-1614.
APA Xu, Liang.,Gao, Li-Min.,Zhao, Jian-Ke.,Liu, Feng.,Zhou, Yan.,...&Zhao, Qing.(2016).Calibration of stray light based on point source transmittance measurement system.guangxue jingmi gongcheng/optics and precision engineering,24(7),1607-1614.
MLA Xu, Liang,et al."Calibration of stray light based on point source transmittance measurement system".guangxue jingmi gongcheng/optics and precision engineering 24.7(2016):1607-1614.

入库方式: OAI收割

来源:西安光学精密机械研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。