Data processing and error analysis for the CE-1 Lunar microwave radiometer
文献类型:期刊论文
作者 | Feng, Jian-Qing![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
刊名 | RESEARCH IN ASTRONOMY AND ASTROPHYSICS
![]() |
出版日期 | 2013-03-01 |
卷号 | 13期号:3页码:359-372 |
关键词 | space vehicles instruments: microwave radiometer Moon: brightness temperature method: data processing error analysis |
英文摘要 | The microwave radiometer (MRM) onboard the Chang' E-1 (CE-1) lunar orbiter is a 4-frequency microwave radiometer, and it is mainly used to obtain the brightness temperature (T-B) of the lunar surface, from which the thickness, temperature, dielectric constant and other related properties of the lunar regolith can be derived. The working mode of the CE-1 MRM, the ground calibration (including the official calibration coefficients), as well as the acquisition and processing of the raw data are introduced. Our data analysis shows that T-B increases with increasing frequency, decreases towards the lunar poles and is significantly affected by solar illumination. Our analysis also reveals that the main uncertainty in T-B comes from ground calibration. |
收录类别 | SCI |
语种 | 英语 |
WOS记录号 | WOS:000316111400011 |
源URL | [http://ir.bao.ac.cn/handle/114a11/5932] ![]() |
专题 | 国家天文台_月球与深空探测研究部 |
作者单位 | Chinese Acad Sci, Natl Astron Observ, Beijing 100012, Peoples R China |
推荐引用方式 GB/T 7714 | Feng, Jian-Qing,Su, Yan,Liu, Jian-Jun,et al. Data processing and error analysis for the CE-1 Lunar microwave radiometer[J]. RESEARCH IN ASTRONOMY AND ASTROPHYSICS,2013,13(3):359-372. |
APA | Feng, Jian-Qing.,Su, Yan.,Liu, Jian-Jun.,Zou, Yong-Liao.,Li, Chun-Lai.,...&Zou Yongliao.(2013).Data processing and error analysis for the CE-1 Lunar microwave radiometer.RESEARCH IN ASTRONOMY AND ASTROPHYSICS,13(3),359-372. |
MLA | Feng, Jian-Qing,et al."Data processing and error analysis for the CE-1 Lunar microwave radiometer".RESEARCH IN ASTRONOMY AND ASTROPHYSICS 13.3(2013):359-372. |
入库方式: OAI收割
来源:国家天文台
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。