The Method of Linear Inflation Control in Ambulatory Blood Pressure Measurement at Finger
文献类型:会议论文
作者 | Pandeng Zhang; Jie Zhang; Quanli Qiu; Ye Chen; Jia Liu |
出版日期 | 2016 |
会议名称 | RCAR2016 |
会议地点 | 吴哥窟,柬埔寨 |
英文摘要 | In 24-hour Non-invasive Blood Pressure (NIBP) measurement, most of the automatic devices use oscillometric method. Accurate measurement of BP at finger using an Oscillometric Waveform (OMW) remains a challenge, particularly when the waveform is collected in inflation, because of the bad signal-to-noise ratio. In the present work, we aimed at measuring Ambulatory Blood Pressure by extracting OMW signal at finger with a new method of linear inflation control and digital signal processing. With the PID control of the pump and real time signal processing, we implemented the blood pressure measurement system and new algorithm. For the validation of our suggested method, we compare the effect of our method and traditional method. Finally, the method is validated in a pilot study against Welch Allyn ProBP 3400 BP monitors. |
收录类别 | 其他 |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/10308] ![]() |
专题 | 深圳先进技术研究院_数字所 |
作者单位 | 2016 |
推荐引用方式 GB/T 7714 | Pandeng Zhang,Jie Zhang,Quanli Qiu,et al. The Method of Linear Inflation Control in Ambulatory Blood Pressure Measurement at Finger[C]. 见:RCAR2016. 吴哥窟,柬埔寨. |
入库方式: OAI收割
来源:深圳先进技术研究院
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。