Continuous and long-term arterial blood pressure monitoring by using h-Shirt
文献类型:会议论文
作者 | Chan, C.H. ; Zhang, Y.T. |
出版日期 | 2008 |
会议名称 | International Conference on Information Technology and Applications in Biomedicine |
会议地点 | Shenzhen |
英文摘要 | Several studies found that the pattern of blood pressure (BP) over a 24-hour period can be a sensitive indicator to predict the outcome and end-organ damage accurately. However, conventional BP meters are unsuitable for measuring BP variation due to their measuring principle, which require an inflatable cuff. To solve this problem, a cuff-less, non-invasive and continuous BP measurement technique are required for 24-hour non-interrupt BP monitoring. Health Shirt (h-Shirt) developed in this study was shown to be one of the possible solution. In this paper, h-Shirt estimation ability was tested against finometer (Finapres Medical Systems, Netherlands) on 10 subjects before, during and after they were riding on a bicycle ergometer. Totally 15 minutes data was captured continuously. The Pulse-Transit-Time capture from h-Shirt correlates with SBP and DBP from finometer by r = -0.77 and -0.51 respectively. In terms of BP value, the overall SBP and DBP difference are 11.29 and 7.15mmHg respectively. |
收录类别 | EI |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/2278] ![]() |
专题 | 深圳先进技术研究院_医工所 |
推荐引用方式 GB/T 7714 | Chan, C.H.,Zhang, Y.T.. Continuous and long-term arterial blood pressure monitoring by using h-Shirt[C]. 见:International Conference on Information Technology and Applications in Biomedicine. Shenzhen. |
入库方式: OAI收割
来源:深圳先进技术研究院
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。