中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Continuous and long-term arterial blood pressure monitoring by using h-Shirt

文献类型:会议论文

作者Chan, C.H. ; Zhang, Y.T.
出版日期2008
会议名称International Conference on Information Technology and Applications in Biomedicine
会议地点Shenzhen
英文摘要Several studies found that the pattern of blood pressure (BP) over a 24-hour period can be a sensitive indicator to predict the outcome and end-organ damage accurately. However, conventional BP meters are unsuitable for measuring BP variation due to their measuring principle, which require an inflatable cuff. To solve this problem, a cuff-less, non-invasive and continuous BP measurement technique are required for 24-hour non-interrupt BP monitoring. Health Shirt (h-Shirt) developed in this study was shown to be one of the possible solution. In this paper, h-Shirt estimation ability was tested against finometer (Finapres Medical Systems, Netherlands) on 10 subjects before, during and after they were riding on a bicycle ergometer. Totally 15 minutes data was captured continuously. The Pulse-Transit-Time capture from h-Shirt correlates with SBP and DBP from finometer by r = -0.77 and -0.51 respectively. In terms of BP value, the overall SBP and DBP difference are 11.29 and 7.15mmHg respectively.
收录类别EI
语种英语
源URL[http://ir.siat.ac.cn:8080/handle/172644/2278]  
专题深圳先进技术研究院_医工所
推荐引用方式
GB/T 7714
Chan, C.H.,Zhang, Y.T.. Continuous and long-term arterial blood pressure monitoring by using h-Shirt[C]. 见:International Conference on Information Technology and Applications in Biomedicine. Shenzhen.

入库方式: OAI收割

来源:深圳先进技术研究院

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。