The study of reconstruction image quality resulting from geometric error in micro-CT system

文献类型:会议论文

作者Dongxing Xi; Jianbao Gui; Jing Zou; Junyan Rong; Zhanli Hu; Xin Liu; Hairong Zheng
出版日期2010
会议名称4th International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2010
会议地点Chengdu, China
英文摘要The reconstruction images often suffer from artifacts caused by misaligned geometry of flat detector in micro-CT. This paper mainly simulates the causality between the reconstruction image quality and the flat detector geometry by the eight-spheres phantom. The simulation results show that the most influential parameter is β (the rotation angle of the detector plane along the center point O) among the five detector parameters, next is a, b, θ and a degressively. We should control β under 7° and a, b offset lower than 8 pixels, when adjusting micro-CT systems. We also show that the relevant correction is efficient to improve the reconstruction image quality affected by the tiny geometric error.