The validation of Echo-PIV technique used in a stenosis model
文献类型:会议论文
作者 | Hecong Zhu; Ming Qian; Yuanwen Zou; Ruibo Song; Niu, Lili; Bo Jiang; Qiu Guang; Hairong Zheng |
出版日期 | 2011 |
会议名称 | 6th International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2011 |
会议地点 | Wuhan, China |
英文摘要 | Echo-PIV technique was used to obtain 2D velocity vectors of flows in a 50% stenosis model. The experiment system was constructed, including a regionally stenotic channel that mimic artery with thrombosis, a fluid circulation system, and an ultrasound imaging equipment. The stenotic channel was fabricated using a pair of male and female plastic rods, a mold, and tissue mimic phantom materials. Ultrasound contrast agents was added into fully developed laminar flow, and Echo-PIV technique was used to obtain the flow velocity pattern in the post-stenosis region of the stenosis model. The simulation result of Fluent software was applied for validation. Echo-PIV technique is confirmed to be a reliable 2D velocimetry technique for blood flow in artery with thrombosis. |
收录类别 | EI |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/3503] ![]() |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2011 |
推荐引用方式 GB/T 7714 | Hecong Zhu,Ming Qian,Yuanwen Zou,et al. The validation of Echo-PIV technique used in a stenosis model[C]. 见:6th International Conference on Bioinformatics and Biomedical Engineering, iCBBE 2011. Wuhan, China. |
入库方式: OAI收割
来源:深圳先进技术研究院
浏览0
下载0
收藏0
其他版本