Poly(Vinyl Alcohol) cryogel phantoms for use in physiological pulsating arterial flow construction and flow dynamics characterization
文献类型:会议论文
作者 | Ming Qian; Ruibo Song; Lili Niu; Weibao Qiu; Hairong Zheng |
出版日期 | 2012 |
会议名称 | 2nd IEEE-EMBS International Conference on Biomedical Engineering and Sciences (IECBES) |
会议地点 | 马来西亚 |
英文摘要 | This paper studied the flow dynamics of a simulated physiological pulsating arterial flow by constructing flow circulation system and performing ultrasound imaging experiments. Polyvinyl Alcohol material was used to fabricate elastic vessel phantom. The mechanical and acoustical properties of the phantoms were characterized. The Young's modulus increases from 19.89 kPa for one cycle to 280.22 kPa for eight cycles. The speed of sound and the sound attenuation increase with free-thaw cycles. The Echopiv measuring results show good parabolic flow profile at two different time periods. The results of this paper could provide basis for following research on how arterial elasticity influence flow dynamics. |
收录类别 | EI |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/4072] ![]() |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2012 |
推荐引用方式 GB/T 7714 | Ming Qian,Ruibo Song,Lili Niu,et al. Poly(Vinyl Alcohol) cryogel phantoms for use in physiological pulsating arterial flow construction and flow dynamics characterization[C]. 见:2nd IEEE-EMBS International Conference on Biomedical Engineering and Sciences (IECBES). 马来西亚. |