Phase-contrast Imaging simulation based on a micro-CT system
文献类型:会议论文
作者 | Jianbao Gui; Jing Zou; Junyan Rong; Zhanli Hu; Qiyang Zhang; Hairong Zheng; Dan Xia |
出版日期 | 2013 |
会议名称 | World Congress on Medical Physics and Biomedical Engineering |
会议地点 | Beijing, China |
英文摘要 | Propagation-based phase-contrast imaging was simulated based on paraxial Fresnel-Kirchoff diffraction integral and spherical wave illumination. Under a developed micro-CT system parameters, the effects of focal-spot size and imaging geometry on phase-contrast imaging have been investigated using a 2 mm thickness polystyrene edge phantom. An equivalent mono-energy was used to substitute the polychromatic spectrum of the micro-focus x-ray source. To consider effects of focal-spot size and detector resolution, the obtained phase-contrast image with an ideal point source was convolved with source intensity distribution and point spread function of detector. Simulations show reasonable influences of the two parameters and are in good agreement with experimental results. |
收录类别 | EI |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/4940] ![]() |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2013 |
推荐引用方式 GB/T 7714 | Jianbao Gui,Jing Zou,Junyan Rong,et al. Phase-contrast Imaging simulation based on a micro-CT system[C]. 见:World Congress on Medical Physics and Biomedical Engineering. Beijing, China. |
入库方式: OAI收割
来源:深圳先进技术研究院
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。