High Frame Rate Ultrasound Displacement Vector Imaging

文献类型:会议论文

作者Kruizinga Pieter; Mastik Frits; Bosch Johannes G; De Jong Nico; Van Soest Gijs; Van Der Steen Antonius F.W.
出版日期2014
会议名称2014 IEEE International Ultrasonics Symposium, IUS 2014
会议地点Chicago, IL, United states
英文摘要The tissue displacement vector obtained with high frame-rate imaging may provide valuable diagnostic information. Measuring 2D displacements caused by rapid dynamic processes, such as arterial pulse waves or shear waves, requires inter-frame displacement precision in the micrometer range. Conventional vector estimation techniques do not satisfy this criterion. In this paper we describe a new robust technique for estimating the 2D displacement vector with the required precision. Echoes reflected from a moving scatterer will cause unique delays along the elements of an array. For each pair of elements the solution for the new scatterer position is exact. By averaging many pair-wise estimates we find a robust estimation of the new scatterer location and thereby the displacement vector. We successfully tested our method on rotating scatterers and on a pulsating carotid artery in transverse view.