中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
In vivo wearable non-invasive glucose monitoring based on dielectric spectroscopy

文献类型:会议论文

作者Yuhang Liu; Meng Xia; Zedong Nie; Jingzhen Li; Yicheng Zeng; Lei Wang
出版日期2016
会议名称ICSP2016
会议地点中国成都
英文摘要Non-invasive glucose monitoring is the most attractive way for people with diabetes. In this paper, we proposed a wearable prototype system for bio-impedance measurement. A tetrapolar Analog-Front-End(4E-AFE) was designed to fully adapt the AD5933 to a four electrode strategy. A normal oral glucose tolerance test (OGTT) in vivo were conducted to investigate the feasibility of the prototype system. The impedance spectroscopy on fingertip within the frequency range of 10 KHz - 100 KHz were measured. The characteristic parameters, R0, R∞, α, fc, Re, Ri, of the bio-impedance equivalent model were extracted. Experimental results shows that the characteristic parameter fc is maximum and then decreasing gradually during period of postprandial 0.5h-1h. Lastly, the value of fc is rising again at postprandial 1.5h, which indicate an inverse behavior that might be an expected response for blood glucose levels based on the OGTT.
收录类别EI
语种英语
源URL[http://ir.siat.ac.cn:8080/handle/172644/10567]  
专题深圳先进技术研究院_医工所
作者单位2016
推荐引用方式
GB/T 7714
Yuhang Liu,Meng Xia,Zedong Nie,et al. In vivo wearable non-invasive glucose monitoring based on dielectric spectroscopy[C]. 见:ICSP2016. 中国成都.

入库方式: OAI收割

来源:深圳先进技术研究院

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。