Dual photometric-conductometric detector for microfluidic chip
文献类型:会议论文
作者 | Shen F(申峰)![]() ![]() ![]() ![]() |
出版日期 | 2009 |
会议名称 | International Conference on Experimental Mechanics 2008, ICEM 2008 |
会议日期 | November 8, 2008 - November 8, 2008 |
会议地点 | Nanjing, China |
关键词 | Analytes Capacitively coupled contactless conductivity detection Conductometrics Detection cells Dual-detectors Electrokinetic Inorganic ions Lab-on-a-chip Laser induced fluorescence Microfluidic chip Rhodamine B Separation efficiency |
通讯作者 | Shen, F. |
中文摘要 | A dual simultaneous photometric-conductometric detector for microfluidic chip is reported. Two different detect methods confocal LIF and moveable C 4D were combined together. They shared a common detection cell and could respond simultaneously. The dual detector offered possibility of measurement one analyte in two different ways and was advantageous in analyses of mixtures containing organic and inorganic ions. To improve the separation efficiency, pinch injection was carried out by a series of electrokinetic manipulations. Rhodamine B was used to evaluate the performance of the dual detector. The dual detector had highly sensitivity and could offer simultaneous information. © 2009 SPIE. |
收录类别 | EI |
会议网址 | 10.1117/12.839069 |
会议录 | Proceedings of SPIE - The International Society for Optical Engineering
![]() |
语种 | 英语 |
源URL | [http://dspace.imech.ac.cn/handle/311007/60465] ![]() |
专题 | 力学研究所_国家微重力实验室 |
通讯作者 | Shen, F. |
推荐引用方式 GB/T 7714 | Shen F,Yu Y,Kang Q,et al. Dual photometric-conductometric detector for microfluidic chip[C]. 见:International Conference on Experimental Mechanics 2008, ICEM 2008. Nanjing, China. November 8, 2008 - November 8, 2008.10.1117/12.839069. |
入库方式: OAI收割
来源:力学研究所