Real-time immunoassay of ferritin using surface plasmon resonance biosensor

文献类型:期刊论文

作者Cui XQ ; Yang F ; Sha YF ; Yang XR
刊名talanta
出版日期2003
卷号60期号:1页码:53-61
关键词BIOSPECIFIC INTERACTION ANALYSIS SERUM FERRITIN IRON IMMOBILIZATION PROTEINS SENSORS SITE
ISSN号0039-9140
通讯作者yang xr
中文摘要a surface plasmon resonance (spr) biosensor was used for the first time to determine the concentration of ferritin in both hbs-ep buffer and serum. the monoclonal antibody was immobilized on the carboxymethyl dextran-modified gold surface by an amine coupling method. the interaction of antibody with antigen was monitored in real-time. the signal was enhanced by sandwich amplification strategy to improve the sensitivity and specificity of the immunoassay, especially in serum. the linear range of the assay in serum is over 30-200 ng ml with the detection limit of 28 ng ml(-1). the sensitivity, specificity, and reproducibility of the assay are satisfactory. the analyte and enhancement antibody-binding surface could be regenerated by ph 2.0 glycine-hcl buffer and the same antibody-immobilized surface could be used for more than 50 cycles of ferritin binding and regeneration.
收录类别SCI
语种英语
WOS记录号WOS:000183183800008
公开日期2010-09-13
源URL[http://202.98.16.49/handle/322003/17893]  
专题长春应用化学研究所_长春应用化学研究所知识产出_期刊论文
推荐引用方式
GB/T 7714
Cui XQ,Yang F,Sha YF,et al. Real-time immunoassay of ferritin using surface plasmon resonance biosensor[J]. talanta,2003,60(1):53-61.
APA Cui XQ,Yang F,Sha YF,&Yang XR.(2003).Real-time immunoassay of ferritin using surface plasmon resonance biosensor.talanta,60(1),53-61.
MLA Cui XQ,et al."Real-time immunoassay of ferritin using surface plasmon resonance biosensor".talanta 60.1(2003):53-61.

入库方式: OAI收割

来源:长春应用化学研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。