Electrochemical H2S sensor with H2SO4 pre-treated Nafion membrane as solid polymer electrolyte

文献类型:期刊论文

作者Yu CB ; Wang YJ ; Hua KF ; Xing W ; Lu TH
刊名sensors and actuators b-chemical
出版日期2002
卷号86期号:2-3页码:259-265
关键词ION-EXCHANGE MEMBRANES HYDROGEN-SULFIDE CONDUCTIVITY
ISSN号0925-4005
通讯作者lu th
中文摘要it is reported for the first time that the performance of the electrochemical h2s sensor with the nation membrane pre-treated with the concentrated h2so4 as the solid electrolyte is much more stable than that for the sensor with the nation membrane without h2so4 pretreatment. the sensitivity of the sensor is about 2.92 mua/ppm. the response time of the sensor is about 9 s. the detection limit is about 0.1 ppm. therefore, this kind of the electrochemical h2s gas sensor may be desirable for the practical application.
收录类别SCI收录期刊论文
语种英语
WOS记录号WOS:000178977500022
公开日期2010-11-03
源URL[http://202.98.16.49/handle/322003/18231]  
专题长春应用化学研究所_长春应用化学研究所知识产出_期刊论文
推荐引用方式
GB/T 7714
Yu CB,Wang YJ,Hua KF,et al. Electrochemical H2S sensor with H2SO4 pre-treated Nafion membrane as solid polymer electrolyte[J]. sensors and actuators b-chemical,2002,86(2-3):259-265.
APA Yu CB,Wang YJ,Hua KF,Xing W,&Lu TH.(2002).Electrochemical H2S sensor with H2SO4 pre-treated Nafion membrane as solid polymer electrolyte.sensors and actuators b-chemical,86(2-3),259-265.
MLA Yu CB,et al."Electrochemical H2S sensor with H2SO4 pre-treated Nafion membrane as solid polymer electrolyte".sensors and actuators b-chemical 86.2-3(2002):259-265.