中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Amplified immunoassay of human IgG using real-time biomolecular interaction analysis (BIA) technology

文献类型:期刊论文

作者Pei RJ ; Cui XQ ; Yang XR ; Wang EK
刊名chinese journal of chemistry
出版日期2002
卷号20期号:5页码:441-446
ISSN号1001-604x
关键词SURFACE-PLASMON RESONANCE BIOSPECIFIC INTERACTION ANALYSIS PROTEIN BIOSENSOR IMMOBILIZATION COMPLEX BINDING SENSORS MATRIX FILMS
通讯作者yang xr
中文摘要an automated biomolecular interaction analysis instrument (bi-acore) based on surface plasmon resonance (spr) has been used to determine human immunoglobulin g (igg) in real time. polyclonal anti-human igg antibody was covalently immobilized to a carboxymethyldextran modified gold film surface. the samples of human igg prepared in hbs buffer were poured over the immobilized surface. the signal amplification antibody was applied to amplify the response signal. after each measurement, the surface was regenerated with 0.1 mol/l h3po4. the assay was rapid, requiring only 30 min for antibody immobilization and 20 min for each subsequent process of immune binding, antibody amplification and regeneration. the antibody immobilized surface had good response to human igg in the range of 0.12-60 nmol/l with a detection limit of 60 pmol/l. the same antibody immobilized surface could be used for more than 110 cycles of binding, amplification and regeneration. the results demonstrate that the sensitivity, specificity and reproducibility of amplified immunoassay using real-time bia technology are satisfactory.
收录类别SCI收录期刊论文
语种英语
WOS记录号WOS:000175713800006
公开日期2010-11-03
源URL[http://202.98.16.49/handle/322003/18603]  
专题长春应用化学研究所_长春应用化学研究所知识产出_期刊论文
推荐引用方式
GB/T 7714
Pei RJ,Cui XQ,Yang XR,et al. Amplified immunoassay of human IgG using real-time biomolecular interaction analysis (BIA) technology[J]. chinese journal of chemistry,2002,20(5):441-446.
APA Pei RJ,Cui XQ,Yang XR,&Wang EK.(2002).Amplified immunoassay of human IgG using real-time biomolecular interaction analysis (BIA) technology.chinese journal of chemistry,20(5),441-446.