Improved spatial calibration for the CXRS system on EAST
文献类型:期刊论文
作者 | Yin, X. H.1,2; Li, Y. Y.2; Fu, J.2; Jiang, D.2,3; Feng, S. Y.1,2; Gu, Y. Q.2; Cheng, Y.2; Lyu, B.2,4; Shi, Y. J.1,5; Ye, M. Y.1 |
刊名 | REVIEW OF SCIENTIFIC INSTRUMENTS |
出版日期 | 2016-11-01 |
卷号 | 87期号:11页码:11E539 |
DOI | 10.1063/1.4962054 |
文献子类 | Article |
英文摘要 | A Charge eXchange Recombination Spectroscopy (CXRS) diagnostic system has been developed to measure profiles of ion temperature and rotation since 2014 on EAST. Several techniques have been developed to improve the spatial calibration of the CXRS diagnostic. The sightline location was obtained by measuring the coordinates of three points on each sightline using an articulated flexible coordinate measuring arm when the vessel was accessible. After vacuum pumping, the effect of pressure change in the vacuum vessel was evaluated by observing the movement of the light spot from back-illuminated sightlines on the first wall using the newly developed articulated inspection arm. In addition, the rotation of the periscope after vacuum pumping was derived by using the Doppler shift of neutral beam emission spectra without magnetic field. Combining these techniques, improved spatial calibration was implemented to provide a complete and accurate description of the EAST CXRS system. Due to the effects of the change of air pressure, alpha similar to 0.4 degrees periscope rotation, yielding a similar to 20 mm movement of the major radius of observation positions to the lower field side, was derived. Results of Zeeman splitting of neutral beam emission spectra with magnetic field also showed good agreement with the calibration results. Published by AIP Publishing. |
WOS研究方向 | Instruments & Instrumentation ; Physics |
语种 | 英语 |
WOS记录号 | WOS:000390242300254 |
源URL | [http://ir.hfcas.ac.cn/handle/334002/21132] |
专题 | 合肥物质科学研究院_中科院等离子体物理研究所 |
作者单位 | 1.Univ Sci & Technol China, Sch Nucl Sci & Technol, Hefei 230026, Peoples R China 2.Chinese Acad Sci, Inst Plasma Phys, Hefei 230031, Peoples R China 3.Anhui Univ, Sch Phys & Mat Sci, Hefei 230601, Peoples R China 4.Chinese Acad Sci, Hefei Sci Ctr, Hefei 230031, Peoples R China 5.Seoul Natl Univ, Dept Nucl Engn, Seoul 151742, South Korea |
推荐引用方式 GB/T 7714 | Yin, X. H.,Li, Y. Y.,Fu, J.,et al. Improved spatial calibration for the CXRS system on EAST[J]. REVIEW OF SCIENTIFIC INSTRUMENTS,2016,87(11):11E539. |
APA | Yin, X. H..,Li, Y. Y..,Fu, J..,Jiang, D..,Feng, S. Y..,...&Wan, B. N..(2016).Improved spatial calibration for the CXRS system on EAST.REVIEW OF SCIENTIFIC INSTRUMENTS,87(11),11E539. |
MLA | Yin, X. H.,et al."Improved spatial calibration for the CXRS system on EAST".REVIEW OF SCIENTIFIC INSTRUMENTS 87.11(2016):11E539. |
入库方式: OAI收割
来源:合肥物质科学研究院
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。