中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Long-term sub second-response monitoring of gaseous ammonia in ambient air by positive inhaling ion mobility spectrometry

文献类型:期刊论文

作者Huang, Wei1; Wang, Weiguo1; Chen, Chuang1; Li, Mei1,3; Peng, Liying1,3; Li, Hang2,3; Liu, Jiwei1,4; Hou, Keyong1; Li, Haiyang1
刊名TALANTA
出版日期2017-12-01
卷号175页码:522-527
关键词Ion Mobility Spectrometry Ammonia Fast Response Sub-second Time Scale Long-term Monitoring
DOI10.1016/j.talanta.2017.07.076
文献子类Article
英文摘要A real-time dynamic measurements of ammonia (NH3) is crucial for understanding the atmospheric nucleation process. A novel method was developed for on line monitoring at the sub-second time scale for the gaseous ammonia in ambient air for months, based on a positive inhaling ion mobility spectrometry (IMS) with a Ni-63 ion source. The selective detection of NH3 was achieved using a high resolution IMS with an optimization of the drift tube temperature above 150 degrees C. This method improved the peak-to-peak resolution significantly, thus avoided the interferences of the adjacent peaks to the quantitative analysis of NH3. The time resolution of the IMS was less than 0.1 s at a data averaging of 10 times. The limit of detection (LOD) achieved at sub-ppb level while a linear response of peak intensity versus concentration of NH3 in the range of 10-60 ppb and 60-400 ppb were obtained. The relative standard deviations (RSD), the confidence level and the errors were 1.06%, 95% and +/- 0.21 ppb by measuring 100 ppb NH3 for 100 times. The effect of ambient humidity could be greatly reduced by using the drift temperature of over 150 degrees C. At last, the application of measuring the NH3 concentration evolutions of Dalian city was performed from June 19 to December 3 in 2015. The results illustrated a potential method of using IMS for a real-time measuring atmospheric NH3 at an unprecedented accuracy and sensitivity with long-term stability.
WOS关键词IONIZATION MASS-SPECTROMETRY ; BRADBURY-NIELSEN-GATE ; ATMOSPHERIC AMMONIA ; ONLINE MEASUREMENT ; NITROGEN-DIOXIDE ; HIGH-RESOLUTION ; BREATH AMMONIA ; AMINES ; GAS ; SPECTROSCOPY
WOS研究方向Chemistry
语种英语
WOS记录号WOS:000412033100070
资助机构Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200) ; Chinese National Key Research and Development Plans(2016YFC0201200)
源URL[http://ir.hfcas.ac.cn:8080/handle/334002/33734]  
专题合肥物质科学研究院_中科院安徽光学精密机械研究所
作者单位1.Chinese Acad Sci, Dalian Inst Chem Phys, Key Lab Separat Sci Analyt Chem, Dalian 116023, Peoples R China
2.Chinese Acad Sci, Key Lab Environm Opt & Technol, Anhui Inst Opt & Fine Mech, Hefei 230031, Peoples R China
3.Univ Chinese Acad Sci, Beijing 100049, Peoples R China
4.Liaoning Normal Univ, Dalian 116029, Peoples R China
推荐引用方式
GB/T 7714
Huang, Wei,Wang, Weiguo,Chen, Chuang,et al. Long-term sub second-response monitoring of gaseous ammonia in ambient air by positive inhaling ion mobility spectrometry[J]. TALANTA,2017,175:522-527.
APA Huang, Wei.,Wang, Weiguo.,Chen, Chuang.,Li, Mei.,Peng, Liying.,...&Li, Haiyang.(2017).Long-term sub second-response monitoring of gaseous ammonia in ambient air by positive inhaling ion mobility spectrometry.TALANTA,175,522-527.
MLA Huang, Wei,et al."Long-term sub second-response monitoring of gaseous ammonia in ambient air by positive inhaling ion mobility spectrometry".TALANTA 175(2017):522-527.

入库方式: OAI收割

来源:合肥物质科学研究院

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。