中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Differentiated correction on the signal intensity dependence of GasBench II-IRMS from blank effect and instrument nonlinear effect

文献类型:期刊论文

作者Wang, J; Wen, XF; Li, SG
刊名INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
出版日期2017
卷号422页码:80-87
关键词GasBench II delta C-13 Signal intensity dependence Blank effect Instrument nonlinear effect
ISSN号1387-3806
英文摘要Continuous-Flow Isotope Ratio Mass Spectrometry (CF-IRMS) coupled with a GasBench II was designed to rapidly analyze delta C-13 in carbonate or CO2 in air. The dependence of measured delta C-13 (delta C-13(m)) on signal intensity (peak area or peak height) was the main error source resulting from blank effect of samples and/or instrument itself effect. To determine the reasons and origins of signal intensity dependence, eight signal intensity gradients for two carbonate (delta C-13 values were -49.64 and 1.61%., respectively) and a gas mixture of 0.3% CO2 in He (delta C-13 = -19.32%.) standards were made respectively. The results showed that there was obvious signal intensity dependence, in particular when sample amount (peak area) was smaller. Given that the dependence was caused by blank effect and/or instrument nonlinear effect, without considering the blank effect, the signal intensity dependence could not be eliminated directly by the instrument nonlinear correction (empirical equation) methods However, it could be eliminated by blank correction (indirect subtraction method). In practice, blank effect should be corrected first, and then the instrument nonlinear effect. (C) 2017 Elsevier B.V. All rights reserved.
学科主题Physics ; Spectroscopy
语种英语
WOS记录号WOS:000417002300010
出版者ELSEVIER SCIENCE BV
源URL[http://ir.igsnrr.ac.cn/handle/311030/43984]  
专题生态系统网络观测与模拟院重点实验室_生态网络实验室
推荐引用方式
GB/T 7714
Wang, J,Wen, XF,Li, SG. Differentiated correction on the signal intensity dependence of GasBench II-IRMS from blank effect and instrument nonlinear effect[J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY,2017,422:80-87.
APA Wang, J,Wen, XF,&Li, SG.(2017).Differentiated correction on the signal intensity dependence of GasBench II-IRMS from blank effect and instrument nonlinear effect.INTERNATIONAL JOURNAL OF MASS SPECTROMETRY,422,80-87.