A calculation method to eliminate gain effect on isotopic measurement using the virtual amplifier multi-collector mass spectrometer

文献类型:期刊论文

作者Chu, Zhuyin; Yang, Yueheng; Qiao, Guangsheng
刊名INTERNATIONAL JOURNAL OF MASS SPECTROMETRY
出版日期2006-06-15
卷号253期号:1-2页码:130-135
关键词calculation method gain factor multi-collector mass spectrometer isotope measurement
ISSN号1387-3806
DOI10.1016/j.ijms.2006.01.034
文献子类Article
英文摘要The virtual amplifier design employed in the ThermoElectron multi-collector mass spectrometers is a novel innovation for high precision isotopic measurement. With the virtual amplifiers, uncertainties of gain factors of amplifiers can be effectively averaged out, so that accuracy and external reproducibility on isotopic measurement can be improved. This paper presents a new calculation method on isotopic measurement using the virtual amplifier mass spectrometers such as ThermoElectron Triton and Neptune. With the calculation method, gain factors of the amplifiers can be completely eliminated out for both raw measured isotopic ratios and normalized isotopic ratios when exponential or power law is used to correct isotope fractionation effect, and therefore, gain calibration is unnecessary even for the high precision measurement purpose. (c) 2006 Elsevier B.V. All rights reserved.
WOS关键词EARLY DIFFERENTIATION ; EARTHS MANTLE ; FRACTIONATION ; ROCKS
WOS研究方向Physics ; Spectroscopy
语种英语
WOS记录号WOS:000238298300016
出版者ELSEVIER SCIENCE BV
源URL[http://ir.iggcas.ac.cn/handle/132A11/67105]  
专题中国科学院地质与地球物理研究所
通讯作者Chu, Zhuyin
作者单位Chinese Acad Sci, Inst Geol & Geophys, State Key Lab Lithospher Evolut, Beijing 100029, Peoples R China
推荐引用方式
GB/T 7714
Chu, Zhuyin,Yang, Yueheng,Qiao, Guangsheng. A calculation method to eliminate gain effect on isotopic measurement using the virtual amplifier multi-collector mass spectrometer[J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY,2006,253(1-2):130-135.
APA Chu, Zhuyin,Yang, Yueheng,&Qiao, Guangsheng.(2006).A calculation method to eliminate gain effect on isotopic measurement using the virtual amplifier multi-collector mass spectrometer.INTERNATIONAL JOURNAL OF MASS SPECTROMETRY,253(1-2),130-135.
MLA Chu, Zhuyin,et al."A calculation method to eliminate gain effect on isotopic measurement using the virtual amplifier multi-collector mass spectrometer".INTERNATIONAL JOURNAL OF MASS SPECTROMETRY 253.1-2(2006):130-135.

入库方式: OAI收割

来源:地质与地球物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。