中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
A QTC-based signal readout for position-sensitive multi-output detectors

文献类型:期刊论文

;
作者Zhou W(周魏); Zhang ZM(章志明); Li DW(李道武); Wang PL(王培林); Feng BT(丰宝桐); Huang XC(黄先超); Hu TT(胡婷婷); Li XH(李晓辉); Chen Y(陈研); Wang YJ(王英杰)
刊名NUCLEAR SCIENCE AND TECHNIQUES ; NUCLEAR SCIENCE AND TECHNIQUES
出版日期2016 ; 2016
卷号27期号:5页码:119
关键词Signal readout Signal readout Charge-to-time conversion Time-to-digital converter H8500 PSPMT Charge-to-time conversion Time-to-digital converter H8500 PSPMT
ISSN号1001-8042
DOI10.1007/s41365-016-0124-4 ; 10.1007/s41365-016-0124-4
通讯作者魏龙 ; 魏龙
文献子类Article
英文摘要A new signal readout method for position-sensitive multi-output detectors, such as those in high-energy spectroscopy measurement and nuclear imaging, was developed by combining the charge division circuit, summing circuit and charge-to-time conversion (QTC) circuit. The 64 outputs of a Hamamatsu H8500 position-sensitive photomultiplier tube were processed, and three digital pulses were generated. The widths of digital pulses were determined using the time-to-digital converter in an field programmable gate array. The energy and position information of incident c-rays is estimated based on the proportionality between the width of digital pulses and input charge created by c-photons. A prototype was built using discrete components and tested, and the energy and position resolutions were improved compared with that obtained with standard ADCs. This method greatly simplifies the front-end electronics and the digital interface. It enables a compact electronics system and an easy integration into an ASIC.; A new signal readout method for position-sensitive multi-output detectors, such as those in high-energy spectroscopy measurement and nuclear imaging, was developed by combining the charge division circuit, summing circuit and charge-to-time conversion (QTC) circuit. The 64 outputs of a Hamamatsu H8500 position-sensitive photomultiplier tube were processed, and three digital pulses were generated. The widths of digital pulses were determined using the time-to-digital converter in an field programmable gate array. The energy and position information of incident c-rays is estimated based on the proportionality between the width of digital pulses and input charge created by c-photons. A prototype was built using discrete components and tested, and the energy and position resolutions were improved compared with that obtained with standard ADCs. This method greatly simplifies the front-end electronics and the digital interface. It enables a compact electronics system and an easy integration into an ASIC.
电子版国际标准刊号2210-3147
WOS关键词PROXIMITY FOCUSING RICH ; GAMMA-CAMERA ; PERFORMANCE EVALUATION ; AEROGEL RADIATOR ; HAMAMATSU H8500 ; PET DETECTOR ; MODULE ; DEPTH ; ARRAY ; PMT
语种英语
WOS记录号WOS:000388510700016 ; WOS:000388510700016
源URL[http://ir.ihep.ac.cn/handle/311005/260540]  
专题高能物理研究所_核技术应用研究中心
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhou W,Zhang ZM,Li DW,et al. A QTC-based signal readout for position-sensitive multi-output detectors, A QTC-based signal readout for position-sensitive multi-output detectors[J]. NUCLEAR SCIENCE AND TECHNIQUES, NUCLEAR SCIENCE AND TECHNIQUES,2016, 2016,27, 27(5):119, 119.
APA 周魏.,章志明.,李道武.,王培林.,丰宝桐.,...&Wei, L.(2016).A QTC-based signal readout for position-sensitive multi-output detectors.NUCLEAR SCIENCE AND TECHNIQUES,27(5),119.
MLA 周魏,et al."A QTC-based signal readout for position-sensitive multi-output detectors".NUCLEAR SCIENCE AND TECHNIQUES 27.5(2016):119.

入库方式: OAI收割

来源:高能物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。