Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF
文献类型:期刊论文
作者 | Zeng, JR; Bian, FG; Wang, J; Li, XH; Wang, YZ; Tian, F; Zhou, P |
刊名 | JOURNAL OF SYNCHROTRON RADIATION
![]() |
出版日期 | 2017 |
卷号 | 24页码:509-520 |
关键词 | Beamline Saxs Absolute Intensity Calibration Molecular Weight |
ISSN号 | 1600-5775 |
DOI | 10.1107/S1600577516019135 |
文献子类 | 期刊论文 |
英文摘要 | The optical system and end-station of bending-magnet beamline BL16B1, dedicated to small-angle X-ray scattering (SAXS) at the Shanghai Synchrotron Radiation Facility, is described. Constructed in 2009 and upgraded in 2013, this beamline has been open to users since May 2009 and supports methodologies including SAXS, wide-angle X-ray scattering (WAXS), simultaneous SAXS/WAXS, grazing-incidence small-angle X-ray scattering (GISAXS) and anomalous small-angle X-ray scattering (ASAXS). Considering that an increasing necessity for absolute calibration of SAXS intensity has been recognized in in-depth investigations, SAXS intensity is re-stated according to the extent of data processing, and the absolute intensity is suggested to be a unified presentation of SAXS data in this article. Theory with a practical procedure for absolute intensity calibration is established based on BL16B1, using glass carbon and water as primary and secondary standards, respectively. The calibration procedure can be completed in minutes and shows good reliability under different conditions. An empirical line of scale factor estimation is also established for any specific SAXS setup at the beamline. Beamline performance on molecular weight (MW) determination is provided as a straightforward application and verification of the absolute intensity calibration. Results show good accuracy with a deviation of less than 10% compared with the known value, which is also the best attainable accuracy in recent studies using SAXS to measure protein MW. Fast MW measurement following the demonstrated method also enables an instant check or pre-diagnosis of the SAXS performance to improve the data acquisition. |
语种 | 英语 |
WOS记录号 | WOS:000395093600019 |
源URL | [http://ir.sinap.ac.cn/handle/331007/27409] ![]() |
专题 | 上海应用物理研究所_中科院上海应用物理研究所2011-2017年 |
推荐引用方式 GB/T 7714 | Zeng, JR,Bian, FG,Wang, J,et al. Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF[J]. JOURNAL OF SYNCHROTRON RADIATION,2017,24:509-520. |
APA | Zeng, JR.,Bian, FG.,Wang, J.,Li, XH.,Wang, YZ.,...&Zhou, P.(2017).Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF.JOURNAL OF SYNCHROTRON RADIATION,24,509-520. |
MLA | Zeng, JR,et al."Performance on absolute scattering intensity calibration and protein molecular weight determination at BL16B1, a dedicated SAXS beamline at SSRF".JOURNAL OF SYNCHROTRON RADIATION 24(2017):509-520. |