The front-end electronics design of dose monitors for beam delivery system of Shanghai Advanced Proton Therapy Facility
文献类型:期刊论文
作者 | Zhao, BQ; Zhao, MH; Liu, M; Yin, CX; Shu, H |
刊名 | NUCLEAR SCIENCE AND TECHNIQUES
![]() |
出版日期 | 2017 |
卷号 | 28期号:6页码:- |
关键词 | Extraction |
ISSN号 | 1001-8042 |
DOI | 10.1007/s41365-017-0230-y |
文献子类 | 期刊论文 |
英文摘要 | A front-end electronics of dose monitor has been developed for measuring irradiation dose to the patient in Shanghai Advanced Proton Therapy Facility. The parallel plate ionization chamber is used for the dose monitoring. Unlike the traditional method of recycling capacitor integration and voltage-to-frequency conversion, this dose monitor electronics uses the trans-impedance amplifier and analog-to-digital conversion method. It performs satisfactorily, with the integral nonlinearity of less than +/- 0.04 nA in the range of -400 to 50 nA and the resolution of about +/- 0.6 nA. |
WOS关键词 | EXTRACTION |
语种 | 英语 |
WOS记录号 | WOS:000402895400011 |
源URL | [http://ir.sinap.ac.cn/handle/331007/28863] ![]() |
专题 | 上海应用物理研究所_中科院上海应用物理研究所2011-2017年 |
推荐引用方式 GB/T 7714 | Zhao, BQ,Zhao, MH,Liu, M,et al. The front-end electronics design of dose monitors for beam delivery system of Shanghai Advanced Proton Therapy Facility[J]. NUCLEAR SCIENCE AND TECHNIQUES,2017,28(6):-. |
APA | Zhao, BQ,Zhao, MH,Liu, M,Yin, CX,&Shu, H.(2017).The front-end electronics design of dose monitors for beam delivery system of Shanghai Advanced Proton Therapy Facility.NUCLEAR SCIENCE AND TECHNIQUES,28(6),-. |
MLA | Zhao, BQ,et al."The front-end electronics design of dose monitors for beam delivery system of Shanghai Advanced Proton Therapy Facility".NUCLEAR SCIENCE AND TECHNIQUES 28.6(2017):-. |
入库方式: OAI收割
来源:上海应用物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。