Performance of a SiPM based semi-monolithic scintillator PET detector

文献类型:期刊论文

作者Xianming Zhang; Xiaohui Wang; Ning Ren; Zhonghua Kuang; Xinhan Deng; Xin Fu; San Wu; Ziru Sang; Zhanli Hu; Dong Liang
刊名PHYSICS IN MEDICINE AND BIOLOGY
出版日期2017
文献子类期刊论文
英文摘要A depth encoding PET detector module using semi-monolithic scintillation crystal single-ended readout by a SiPM array was built and its performance was measured. The semi-monolithic scintillator detector consists of 11 polished LYSO slices measuring 1 × 11.6 × 10 mm3. The slices are glued together with enhanced specular reflector (ESR) in between and outside of the slices. The bottom surface of the slices is coupled to a 4 × 4 SiPM array with a 1 mm light guide and silicon grease between them. No reflector is used on the top surface and two sides of the slices to reduce the scintillation photon reflection. The signals of the 4 × 4 SiPM array are grouped along rows and columns separately into eight signals. Four SiPM column signals are used to identify the slices according to the center of the gravity of the scintillation photon distribution in the pixelated direction. Four SiPM row signals are used to estimate the y (monolithic direction) and z (depth of interaction) positions according to the center of the gravity and the width of the scintillation photon distribution in the monolithic direction, respectively. The detector was measured with 1 mm sampling interval in both the y and z directions with electronic collimation by using a 0.25 mm diameter 22Na point source and a 1 × 1 × 20 mm3 LYSO crystal detector. An average slice based energy resolution of 14.9% was obtained. All slices of 1 mm thick were clearly resolved and a detector with even thinner slices could be used. The y positions calculated with the center of gravity method are different for interactions happening at the same y, but different z positions due to depth dependent edge effects. The least-square minimization and the maximum likelihood positioning algorithms were developed and both methods improved the spatial resolution at the edges of the detector as compared with the center of gravity method. A mean absolute error (MAE) which is defined as the probability-weighted mean of the absolute value of the positioning error is used to evaluate the spatial resolution. An average MAE spatial resolution of ~1.15 mm was obtained in both y and z directions without rejection of the multiple scattering events. The average MAE spatial resolution was ~0.7 mm in both y and z directions after the multiple scattering events were rejected. The timing resolution of the detector is 575 ps. In the next step, long rectangle detector will be built to reduce edge effects and improve the spatial resolution of the semi-monolithic detector. Thick detector up to 20 mm will be explored and the positioning algorithms will be further optimized.
URL标识查看原文
语种英语
源URL[http://ir.siat.ac.cn:8080/handle/172644/12104]  
专题深圳先进技术研究院_医工所
作者单位PHYSICS IN MEDICINE AND BIOLOGY
推荐引用方式
GB/T 7714
Xianming Zhang,Xiaohui Wang,Ning Ren,et al. Performance of a SiPM based semi-monolithic scintillator PET detector[J]. PHYSICS IN MEDICINE AND BIOLOGY,2017.
APA Xianming Zhang.,Xiaohui Wang.,Ning Ren.,Zhonghua Kuang.,Xinhan Deng.,...&Yongfeng Yang.(2017).Performance of a SiPM based semi-monolithic scintillator PET detector.PHYSICS IN MEDICINE AND BIOLOGY.
MLA Xianming Zhang,et al."Performance of a SiPM based semi-monolithic scintillator PET detector".PHYSICS IN MEDICINE AND BIOLOGY (2017).

入库方式: OAI收割

来源:深圳先进技术研究院

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。