Design of 5-channel On-coil Shimming Coil for Rat Brain MRI Guided High Intensity Focused Ultrasound: a preliminary study
文献类型:会议论文
作者 | Jo Lee; Jianhong Wen; Chao Zou; Xiaoqing Hu; Xiaoliang Zhang; Xin Liu; Ye Li |
出版日期 | 2017 |
会议日期 | 2017 |
会议地点 | Hawaii,USA |
英文摘要 | Signal-to-noise ratio (SNR) and homogeneity of the static magnetic ሬeld B0 are critically important in achieving accurate temperature measurement for MRI guided HIFU. In this work, a 5-channel shimming coil on the top of a 3-channel RF coil array at 3T has been fabricated to increase the B0 homogeneity and the SNR in region of interest. Phantom studies have demonstrated the the capability of the proposed shimming coil to improve B0 homogeneity and SNR of the images, which will lead to accurate temperature mapping |
语种 | 英语 |
源URL | [http://ir.siat.ac.cn:8080/handle/172644/12234] ![]() |
专题 | 深圳先进技术研究院_医工所 |
作者单位 | 2017 |
推荐引用方式 GB/T 7714 | Jo Lee,Jianhong Wen,Chao Zou,et al. Design of 5-channel On-coil Shimming Coil for Rat Brain MRI Guided High Intensity Focused Ultrasound: a preliminary study[C]. 见:. Hawaii,USA. 2017. |