Optical design laboratory test and calibration of airborne long wave infrared imaging spectrometer
文献类型:期刊论文
作者 | Yuan LY; He ZP; Lv G; Wang YM; Li CL; Xie JN; Wang JY |
刊名 | OPTICS EXPRESS |
出版日期 | 2017 |
卷号 | 25期号:19页码:22440-22454 |
DOI | 10.1364/OE.25.022440 |
英文摘要 | We discuss and evaluate a long wave infrared imaging spectrometer in terms of its opto-mechanical design and analysis alignment testing and calibration. The instrument is a practical airborne sensor achieving high spectral resolution and sensitive noise equivalent delta temperature. The instrument operates in the 8 to 12.5 mu m spectral region with 28.85 nm spectral sampling 1 mrad instantaneous field of view and >40 degrees cross track field. The instrument comprises three uniform sub-modules with identical design parameters and performances. The sub-module design is based on a refractive foreoptics feeding an allreflective spectrometer. The optical form of the spectrometer is a double-pass reflective triplet with a flat grating which has a fast f/2 and high optical throughput. Cryogenic optics of 100 K is implemented only for the spectrometer. Assembly and thermal deformation and focusing adjustment design are particularly considered for this low temperature. All the mirrors of the spectrometer are opto-mechanical-integrated designed and manufactured by single-point diamond turning technology. We consider the center sub-module as an example and we present its laboratory testing results and calibration;the results indicate the instrument's potential value in airborne sensing. (C) 2017 Optical Society of America |
源URL | [http://202.127.2.71:8080/handle/181331/12266] |
专题 | 上海技术物理研究所_上海技物所 |
作者单位 | Shanghai Inst Tech Phys |
推荐引用方式 GB/T 7714 | Yuan LY,He ZP,Lv G,et al. Optical design laboratory test and calibration of airborne long wave infrared imaging spectrometer[J]. OPTICS EXPRESS,2017,25(19):22440-22454. |
APA | Yuan LY.,He ZP.,Lv G.,Wang YM.,Li CL.,...&Wang JY.(2017).Optical design laboratory test and calibration of airborne long wave infrared imaging spectrometer.OPTICS EXPRESS,25(19),22440-22454. |
MLA | Yuan LY,et al."Optical design laboratory test and calibration of airborne long wave infrared imaging spectrometer".OPTICS EXPRESS 25.19(2017):22440-22454. |
入库方式: OAI收割
来源:上海技术物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。