A devices for measuring low temperature of filament in ion source of VG 354 thermal ionization mass spectrometer and its application
文献类型:期刊论文
作者 | Xiao, YK; Wei, HZ; Wang, QZ; Zhang, CG; Sun, AD |
刊名 | CHINESE JOURNAL OF ANALYTICAL CHEMISTRY
![]() |
出版日期 | 2002-10-01 |
卷号 | 30期号:10页码:1272-1276 |
关键词 | Mass Spectrometry Thermal Ion emissIon Measurement Of Temperature |
文献子类 | Article |
英文摘要 | A device was designed to measure low temperature of filament in ion source of VG 354 thermal ionization mass spectrometry by assembling thermocouple thermometer and connector. This device can he used to measure instantly temperature on the surface of single filament in ion source. The functional correlation between the measured temperature of filament and the heating current through the filament was characterized by aft equation of y = -ax(3) + bx(2) - cx + d. The error of measured temperature is less than +/- 2 degreesC. The device was also applied to study the characteristics of non-reductive thermal ion emission of graphite in thermal ionization mass spectrometry and to study mechanism of M2BO2+ and M2X+ emission in the presence of graphite. |
WOS关键词 | PRECISION ISOTOPIC MEASUREMENT ; EMISSION |
WOS研究方向 | Chemistry |
语种 | 英语 |
WOS记录号 | WOS:000178919300029 |
源URL | [http://ir.isl.ac.cn/handle/363002/6071] ![]() |
专题 | 青海盐湖研究所_青海盐湖研究所知识仓储 中国科学院青海盐湖研究所 |
作者单位 | Chinese Acad Sci, Qinghai Inst Salt Lakes, Xining 810008, Peoples R China |
推荐引用方式 GB/T 7714 | Xiao, YK,Wei, HZ,Wang, QZ,et al. A devices for measuring low temperature of filament in ion source of VG 354 thermal ionization mass spectrometer and its application[J]. CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,2002,30(10):1272-1276. |
APA | Xiao, YK,Wei, HZ,Wang, QZ,Zhang, CG,&Sun, AD.(2002).A devices for measuring low temperature of filament in ion source of VG 354 thermal ionization mass spectrometer and its application.CHINESE JOURNAL OF ANALYTICAL CHEMISTRY,30(10),1272-1276. |
MLA | Xiao, YK,et al."A devices for measuring low temperature of filament in ion source of VG 354 thermal ionization mass spectrometer and its application".CHINESE JOURNAL OF ANALYTICAL CHEMISTRY 30.10(2002):1272-1276. |
入库方式: OAI收割