中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes

文献类型:期刊论文

作者Luo, Yong; Wu, Dapeng; Zeng, Shaojiang; Gai, Hongwei; Long, Zhicheng; Shen, Zheng; Dai, Zhongpeng; Qin, Jianhua; Lin, Bingcheng
刊名analytical chemistry
出版日期2006-09-01
卷号78期号:17页码:6074-6080
产权排序1;1
通讯作者秦建华 ; 林炳承
英文摘要a novel sample injection method for chip ce was presented. this injection method uses hydrostatic pressure, generated by emptying the sample waste reservoir, for sample loading and electrokinetic force for dispensing. the injection was performed on a double-cross microchip. one cross, created by the sample and separation channels, is used for formation of a sample plug. another cross, formed by the sample and controlling channels, is used for plug control. by varying the electric field in the controlling channel, the sample plug volume can be linearly adjusted. hydrostatic pressure takes advantage of its ease of generation on a microfluidic chip, without any electrode or external pressure pump, thus allowing a sample injection with a minimum number of electrodes. the potential of this injection method was demonstrated by a four-separation-channel chip ce system. in this system, parallel sample separation can be achieved with only two electrodes, which is otherwise impossible with conventional injection methods. hydrostatic pressure maintains the sample composition during the sample loading, allowing the injection to be free of injection bias.
WOS标题词science & technology ; physical sciences
类目[WOS]chemistry, analytical
研究领域[WOS]chemistry
关键词[WOS]capillary electrophoresis ; microchip electrophoresis ; microfluidic chips ; hydrodynamic injection ; flow ; performance ; system ; devices ; separations
收录类别SCI
原文出处true
语种英语
WOS记录号WOS:000240158000024
公开日期2010-11-30
源URL[http://159.226.238.44/handle/321008/97221]  
专题大连化学物理研究所_中国科学院大连化学物理研究所
作者单位Chinese Acad Sci, Dalian Inst Chem Phys, Lab Microfluid Syst Biol, Dalian, Liaoning Prov, Peoples R China
推荐引用方式
GB/T 7714
Luo, Yong,Wu, Dapeng,Zeng, Shaojiang,et al. Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes[J]. analytical chemistry,2006,78(17):6074-6080.
APA Luo, Yong.,Wu, Dapeng.,Zeng, Shaojiang.,Gai, Hongwei.,Long, Zhicheng.,...&Lin, Bingcheng.(2006).Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes.analytical chemistry,78(17),6074-6080.
MLA Luo, Yong,et al."Double-cross hydrostatic pressure sample injection for chip CE: Variable sample plug volume and minimum number of electrodes".analytical chemistry 78.17(2006):6074-6080.

入库方式: OAI收割

来源:大连化学物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。