A novel charge sensitive preamplifier without the feedback resistor

文献类型:期刊论文

作者BERTUCCIO, G; REHAK, P; XI, DM
刊名Nuclear instruments & methods in physics research section a-accelerators spectrometers detectors and associated equipment
出版日期1993-03-01
卷号326期号:1-2页码:71-76
ISSN号0168-9002
通讯作者Bertuccio, g()
英文摘要A novel charge sensitive preamplifier#1 which has no resistor in parallel with the feedback capacitor is presented. no external device or circuit is required to discharge the feedback capacitor. the detector leakage and signal current flows away through the gate of the input jfet which works with its gate to source junction slightly forward biased. the dc stabilization of the preamplifier is accomplished by an additional feedback loop, which permits to equalize the current flowing through the forward biased gate to source junction and the current coming from the detector. an equivalent noise charge of less than 20 electrons rms has been measured at room temperature by using an input jfet with a transconductance to gate capacitance ratio of 4 ms/5.4 pf.
WOS研究方向Instruments & Instrumentation ; Nuclear Science & Technology ; Physics ; Spectroscopy
WOS类目Instruments & Instrumentation ; Nuclear Science & Technology ; Physics, Particles & Fields ; Spectroscopy
语种英语
WOS记录号WOS:A1993KP97200011
出版者ELSEVIER SCIENCE BV
URI标识http://www.irgrid.ac.cn/handle/1471x/2175966
专题高能物理研究所
通讯作者BERTUCCIO, G
作者单位1.BROOKHAVEN NATL LAB,UPTON,NY 11973
2.INST HIGH ENERGY PHYS,BEIJING,PEOPLES R CHINA
推荐引用方式
GB/T 7714
BERTUCCIO, G,REHAK, P,XI, DM. A novel charge sensitive preamplifier without the feedback resistor[J]. Nuclear instruments & methods in physics research section a-accelerators spectrometers detectors and associated equipment,1993,326(1-2):71-76.
APA BERTUCCIO, G,REHAK, P,&XI, DM.(1993).A novel charge sensitive preamplifier without the feedback resistor.Nuclear instruments & methods in physics research section a-accelerators spectrometers detectors and associated equipment,326(1-2),71-76.
MLA BERTUCCIO, G,et al."A novel charge sensitive preamplifier without the feedback resistor".Nuclear instruments & methods in physics research section a-accelerators spectrometers detectors and associated equipment 326.1-2(1993):71-76.