A correction method of the anode wire modulation for 2d mwpcs detector
文献类型:期刊论文
作者 | Wen, Z. W.1,2,3; Qi, H. R.2,3; Zhang, Y. L.1,2,3; Wang, H. Y.2,3,4; Liu, L.1,2,3; Li, Y. H.1 |
刊名 | Journal of instrumentation
![]() |
出版日期 | 2018-04-01 |
卷号 | 13页码:10 |
关键词 | Wire chambers (mwpc thin-gap chambers drift chambers Drift tubes proportional chambers etc) Gaseous detectors Neutron detectors (cold thermal fast neutrons) X-ray detectors |
ISSN号 | 1748-0221 |
DOI | 10.1088/1748-0221/13/04/t04008 |
通讯作者 | Qi, h. r.(qihr@ihep.ac.cn) |
英文摘要 | The linearity performance of 2d multi-wire proportional chambers (mwpcs) detector across the anode wires is modulated by the discrete anode wires. a mwpcs dectector with the 2 mm anode wire spacing was developed to study the anode wire modulation effect. the 2d lineartity performance was measured with a fe-55 source which was moved by a electric mobile platform. the experimental results show that the deviation of the measured position depends upon the incident position in the axis across the anode wires and the curve between the measured position and the incident position is consistent with the sine function whose period is equal to the anode wire spacing. a correction method of the measured position across the anode wire direction was obtained by fitting the curve between the measured position and the incident position. the non-linearity of the measured position across the anode wire direction is reduced about 0.085% and the imaging capability is obviously improved after the data is modified by the correction method. |
WOS关键词 | NEUTRON DETECTOR ; PROPORTIONAL CHAMBERS ; PERFORMANCE |
WOS研究方向 | Instruments & Instrumentation |
WOS类目 | Instruments & Instrumentation |
语种 | 英语 |
WOS记录号 | WOS:000431024000003 |
出版者 | IOP PUBLISHING LTD |
URI标识 | http://www.irgrid.ac.cn/handle/1471x/2177590 |
专题 | 高能物理研究所 |
通讯作者 | Qi, H. R. |
作者单位 | 1.Lanzhou Univ, Lanzhou 730000, Gansu, Peoples R China 2.State Key Lab Particle Detect & Elect, Beijing 100049, Peoples R China 3.Chinese Acad Sci, Inst High Energy Phys, Beijing 100049, Peoples R China 4.Univ Chinese Acad Sci, Beijing 100049, Peoples R China |
推荐引用方式 GB/T 7714 | Wen, Z. W.,Qi, H. R.,Zhang, Y. L.,et al. A correction method of the anode wire modulation for 2d mwpcs detector[J]. Journal of instrumentation,2018,13:10. |
APA | Wen, Z. W.,Qi, H. R.,Zhang, Y. L.,Wang, H. Y.,Liu, L.,&Li, Y. H..(2018).A correction method of the anode wire modulation for 2d mwpcs detector.Journal of instrumentation,13,10. |
MLA | Wen, Z. W.,et al."A correction method of the anode wire modulation for 2d mwpcs detector".Journal of instrumentation 13(2018):10. |
入库方式: iSwitch采集
来源:高能物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。