Study on temperature coefficient of cdte detector used for x-rays detection

文献类型:期刊论文

作者Guo, Siming1; Wu, Jinjie1; Du, Haiyan1; Zhang, Jian1; Li, Xufang2; Liu, Congzhan2; Zhang, Shuai1; Li, Chengze1; Huo, Binbin1; Liao, Zhenyu1
刊名Nuclear technology & radiation protection
出版日期2017-09-01
卷号32期号:3页码:256-260
关键词Cdte detector Temperature coefficient X-ray Detection efficiency Energy resolution
ISSN号1451-3994
DOI10.2298/ntrp1703256g
通讯作者Guo, siming(gsm@nim.ac.cn)
英文摘要The temperature of the working environment is one of the key factors in determining the properties of semiconductor detectors, and it affects the absolute accuracy and stability of the standard detector. in order to determine the temperature coefficient of cdte detector used for x-rays detection, a precise temperature control system was designed. in this experiment, detectors and radiographic source were set inside a thermostat with temperature of 0 similar to 40 degrees c, so that the temperature can be regulated for the test of the temperature coefficient of cdte detector. studies had shown that, with the increase of the temperature, the energy resolution and detection efficiency of the cdte detector would deteriorate, and under 10 degrees c the detectors have better performance with the 8 kev x-rays.
WOS研究方向Nuclear Science & Technology
WOS类目Nuclear Science & Technology
语种英语
WOS记录号WOS:000416298800010
出版者VINCA INST NUCLEAR SCI
URI标识http://www.irgrid.ac.cn/handle/1471x/2177791
专题高能物理研究所
通讯作者Guo, Siming
作者单位1.Natl Inst Metrol, Beijing, Peoples R China
2.Inst High Energy Phys, Beijing, Peoples R China
推荐引用方式
GB/T 7714
Guo, Siming,Wu, Jinjie,Du, Haiyan,et al. Study on temperature coefficient of cdte detector used for x-rays detection[J]. Nuclear technology & radiation protection,2017,32(3):256-260.
APA Guo, Siming.,Wu, Jinjie.,Du, Haiyan.,Zhang, Jian.,Li, Xufang.,...&Liu, Ying.(2017).Study on temperature coefficient of cdte detector used for x-rays detection.Nuclear technology & radiation protection,32(3),256-260.
MLA Guo, Siming,et al."Study on temperature coefficient of cdte detector used for x-rays detection".Nuclear technology & radiation protection 32.3(2017):256-260.

入库方式: iSwitch采集

来源:高能物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。