Study on temperature coefficient of cdte detector used for x-rays detection
文献类型:期刊论文
作者 | Guo, Siming1; Wu, Jinjie1; Du, Haiyan1; Zhang, Jian1; Li, Xufang2; Liu, Congzhan2; Zhang, Shuai1; Li, Chengze1; Huo, Binbin1; Liao, Zhenyu1 |
刊名 | Nuclear technology & radiation protection
![]() |
出版日期 | 2017-09-01 |
卷号 | 32期号:3页码:256-260 |
关键词 | Cdte detector Temperature coefficient X-ray Detection efficiency Energy resolution |
ISSN号 | 1451-3994 |
DOI | 10.2298/ntrp1703256g |
通讯作者 | Guo, siming(gsm@nim.ac.cn) |
英文摘要 | The temperature of the working environment is one of the key factors in determining the properties of semiconductor detectors, and it affects the absolute accuracy and stability of the standard detector. in order to determine the temperature coefficient of cdte detector used for x-rays detection, a precise temperature control system was designed. in this experiment, detectors and radiographic source were set inside a thermostat with temperature of 0 similar to 40 degrees c, so that the temperature can be regulated for the test of the temperature coefficient of cdte detector. studies had shown that, with the increase of the temperature, the energy resolution and detection efficiency of the cdte detector would deteriorate, and under 10 degrees c the detectors have better performance with the 8 kev x-rays. |
WOS研究方向 | Nuclear Science & Technology |
WOS类目 | Nuclear Science & Technology |
语种 | 英语 |
WOS记录号 | WOS:000416298800010 |
出版者 | VINCA INST NUCLEAR SCI |
URI标识 | http://www.irgrid.ac.cn/handle/1471x/2177791 |
专题 | 高能物理研究所 |
通讯作者 | Guo, Siming |
作者单位 | 1.Natl Inst Metrol, Beijing, Peoples R China 2.Inst High Energy Phys, Beijing, Peoples R China |
推荐引用方式 GB/T 7714 | Guo, Siming,Wu, Jinjie,Du, Haiyan,et al. Study on temperature coefficient of cdte detector used for x-rays detection[J]. Nuclear technology & radiation protection,2017,32(3):256-260. |
APA | Guo, Siming.,Wu, Jinjie.,Du, Haiyan.,Zhang, Jian.,Li, Xufang.,...&Liu, Ying.(2017).Study on temperature coefficient of cdte detector used for x-rays detection.Nuclear technology & radiation protection,32(3),256-260. |
MLA | Guo, Siming,et al."Study on temperature coefficient of cdte detector used for x-rays detection".Nuclear technology & radiation protection 32.3(2017):256-260. |
入库方式: iSwitch采集
来源:高能物理研究所
浏览0