中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
A 22-Channel RF coil array for fetus MR imaging at 3T

文献类型:会议论文

作者Chao Luo; Guoxi Xie; Jo Lee; Xing Yang; Xiaoliang Zhang; Xin Liu; Ye Li
出版日期2018
会议日期2018
会议地点Paris, France
英文摘要Due to lack of dedicated fetal imaging RF coils, the system body coil is often used to acquire fetal images. This setup is not optimized and offers limited sensitivity and image quality. In this work, we designed and manufactured a 22-channel flexible coil array for fetal examinations. Compared with Siemens 6-channel body coil, the proposed fetal coil array achieves significant improvements in imaging coverage, image SNR and parallel acceleration capability.
URL标识查看原文
源URL[http://ir.siat.ac.cn:8080/handle/172644/14535]  
专题深圳先进技术研究院_医工所
推荐引用方式
GB/T 7714
Chao Luo,Guoxi Xie,Jo Lee,et al. A 22-Channel RF coil array for fetus MR imaging at 3T[C]. 见:. Paris, France. 2018.

入库方式: OAI收割

来源:深圳先进技术研究院

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。