Simultaneous displacement and T2 mapping of High-intensity focused ultrasound therapy

文献类型:会议论文

作者Qiao Yangzi; Zou Chao; Xu Zongwei; Cheng Chuanli; Wan Qian; Tie Changjun; Liu Xin; Zheng Hairong
出版日期2018
会议日期2018
会议地点Paris Expo Porte de Versailles Paris, France
英文摘要In this work, a hybrid ARFI sequence based on segmented SE-EPI is proposed to simultaneously monitor the displacement and T2 change of tissue during HIFU therapy. The reliability of this sequence was validated first. The quantified displacement and T2 show good consistence with the reference ARFI and SE results. The hybrid sequence was then applied before and after HIFU therapy to evaluate the treatment effects. With the occurrence of ablative lesion, T2 relaxation time decreased in the lesion center and increased in the boundary. While the displaced region (region with obvious displacement) and the maximal displacement at focus both enlarged. In general, this hybrid ARFI is a potentially useful HIFU monitoring method in clinical application.
URL标识查看原文
源URL[http://ir.siat.ac.cn:8080/handle/172644/14548]  
专题深圳先进技术研究院_医工所
推荐引用方式
GB/T 7714
Qiao Yangzi,Zou Chao,Xu Zongwei,et al. Simultaneous displacement and T2 mapping of High-intensity focused ultrasound therapy[C]. 见:. Paris Expo Porte de Versailles Paris, France. 2018.