Fiber-optic fluorescence sensor for dissolved oxygen detection based on fluorinated xerogel immobilized with ruthenium (II) complex

文献类型:期刊论文

作者Xiong, Yan; Xu, Jing; Zhu, Dao-qian; Duan, Chun-feng; Guan, Ya-feng
刊名journal of sol-gel science and technology
出版日期2010-02-01
卷号53期号:2页码:441-447
关键词Fiber-optic sensor Dissolved oxygen detection Two-fiber probe Fluorinated xerogel Ruthenium (II) complex
产权排序1;1
通讯作者关亚风
英文摘要a fiber-optic sensor based on fluorescence quenching was designed for dissolved oxygen (do) detection. the fluorinated xerogel-based sensing film of the present sensor was prepared from 3, 3, 3-trifluoropropyltrimethoxysilane (tfp-trimos). oxygen-sensitive fluorophores of tris (2, 2'- bipyridine) ruthenium (ii) (ru(bpy) (3) (2+) ) were immobilized in the sensing film and the emission fluorescence was quenched by dissolved oxygen. in the sensor fabrication, a two-fiber probe was employed to obtain the best fluorescence collection efficiency and the sensing film was attached to the probe end. scanning electron microscope (sem), uv-vis absorption spectroscopy (uv-vis) and fourier transform infrared spectroscopy (ftir) measurements have been used to characterize the sensing film. the sensor sensitivity is quantified by i (deoxy)/i (oxy), where i (deoxy) and i (oxy) represented the detected fluorescence intensities in fully deoxygenated and fully oxygenated environments, respectively. compared with tetramethoxysilane (tmos) and methyltriethoxysilane (mtms)-derived sensing films, tfp-trimos-based sensor exhibited excellent performances in dissolved oxygen detection with short response time of 4 s, low limit of detection (lod) of 0.04 ppm (r.s.d. = 2.5%), linear stern-volmer calibration plot from 0 to 40 ppm and long-term stability during the past 10 months. the reasons for the preferable performances of tfp-trimos-based sensing film were discussed.
WOS标题词science & technology ; technology
类目[WOS]materials science, ceramics
研究领域[WOS]materials science
关键词[WOS]trifluoroethylmethacrylate copolymer film ; luminescence change ; optical sensor ; thin-films ; matrix
收录类别SCI
原文出处false
语种英语
WOS记录号WOS:000274318400043
公开日期2010-11-30
源URL[http://159.226.238.44/handle/321008/103419]  
专题大连化学物理研究所_中国科学院大连化学物理研究所
作者单位Chinese Acad Sci, Dalian Inst Chem Phys, Dept Instrumentat & Analyt Chem, Key Lab Separat & Anal, Dalian 116023, Peoples R China
推荐引用方式
GB/T 7714
Xiong, Yan,Xu, Jing,Zhu, Dao-qian,et al. Fiber-optic fluorescence sensor for dissolved oxygen detection based on fluorinated xerogel immobilized with ruthenium (II) complex[J]. journal of sol-gel science and technology,2010,53(2):441-447.
APA Xiong, Yan,Xu, Jing,Zhu, Dao-qian,Duan, Chun-feng,&Guan, Ya-feng.(2010).Fiber-optic fluorescence sensor for dissolved oxygen detection based on fluorinated xerogel immobilized with ruthenium (II) complex.journal of sol-gel science and technology,53(2),441-447.
MLA Xiong, Yan,et al."Fiber-optic fluorescence sensor for dissolved oxygen detection based on fluorinated xerogel immobilized with ruthenium (II) complex".journal of sol-gel science and technology 53.2(2010):441-447.

入库方式: OAI收割

来源:大连化学物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。