中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
The development of charge detection-quadrupole ion trap mass spectrometry driven by rectangular and triangular waves

文献类型:期刊论文

作者Xiong, Caiqiao1,2; Xu, Gaoping3; Zhou, Xiaoyu1,2; Wang, Jianing1,2; Tang, Yin1,2; Chen, Rui1,2; Peng, Wen-Ping4; Chang, Huan-Cheng5; Nie, Zongxiu1,2,6
刊名ANALYST
出版日期2012
卷号137期号:5页码:1199-1204
ISSN号0003-2654
DOI10.1039/c2an15756j
英文摘要In this article, the charge detection quadrupole mass spectrometry (CD-ITMS) driven by rectangular and triangular waveforms (rect-CD-ITMS and tri-CD-ITMS) was developed for the characterization of microparticles. Since the frequency scan of rectangular and triangular waveform could be realized easier than that of sinusoidal waveform, this research intends to provide simpler operation modes for CD-ITMS. In order to demonstrate the feasibility of rect-CD-ITMS and tri-CD-ITMS, the discharge onset voltage, ejection point of analyzed particles, and the achieved mass resolution were analyzed and compared with the case in conventional sinusoidal CD-ITMS (sin-CD-ITMS). The results indicated that the rect-CD-ITMS and tri-CD-ITMS can work well for the mass measurement of microparticles by using frequency scan. Identical mass resolutions were achieved under the same root mean square (RMS) voltage of different waveforms. The mass resolution was further improved by increasing the applied voltage and signal-to-noise ratio (S/N) of charge detector. Moreover, the rect-CD-ITMS and tri-CD-ITMS were applied to characterize red blood cells (RBCs). According to the obtained mean masses and mass distributions, normal and anemic RBCs were distinguished successfully.
语种英语
WOS记录号WOS:000300038400021
出版者ROYAL SOC CHEMISTRY
源URL[http://ir.iccas.ac.cn/handle/121111/48259]  
专题中国科学院化学研究所
通讯作者Nie, Zongxiu
作者单位1.Chinese Acad Sci, Inst Chem, Key Lab Analyt Chem Living Biosyst, Beijing 100190, Peoples R China
2.Beijing Natl Lab Mol Sci, Beijing 100190, Peoples R China
3.Jiujiang Univ, Sch Sci, Jiujiang 332005, Peoples R China
4.Natl Dong Hwa Univ, Dept Phys, Shoufeng 97401, Hualien, Taiwan
5.Acad Sinica, Inst Atom & Mol Sci, Taipei 106, Taiwan
6.Beijing Ctr Mass Spectrometry, Beijing 100190, Peoples R China
推荐引用方式
GB/T 7714
Xiong, Caiqiao,Xu, Gaoping,Zhou, Xiaoyu,et al. The development of charge detection-quadrupole ion trap mass spectrometry driven by rectangular and triangular waves[J]. ANALYST,2012,137(5):1199-1204.
APA Xiong, Caiqiao.,Xu, Gaoping.,Zhou, Xiaoyu.,Wang, Jianing.,Tang, Yin.,...&Nie, Zongxiu.(2012).The development of charge detection-quadrupole ion trap mass spectrometry driven by rectangular and triangular waves.ANALYST,137(5),1199-1204.
MLA Xiong, Caiqiao,et al."The development of charge detection-quadrupole ion trap mass spectrometry driven by rectangular and triangular waves".ANALYST 137.5(2012):1199-1204.

入库方式: OAI收割

来源:化学研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。