Instrument compliance effects revisited: linear viscoelastic measurements
文献类型:期刊论文
作者 | Liu, Chen-Yang1,2; Yao, Minglong3; Garritano, Ronald G.3; Franck, Aloyse J.4; Bailly, Christian2 |
刊名 | RHEOLOGICA ACTA
![]() |
出版日期 | 2011-06-01 |
卷号 | 50期号:5-6页码:537-546 |
关键词 | Compliance Dynamic Moduli Viscoelasticity Rheometer |
ISSN号 | 0035-4511 |
DOI | 10.1007/s00397-011-0560-3 |
英文摘要 | Torsional compliance of the torque transducer can be an important issue in linear viscoelastic (LVE) measurements when the sample stiffness is high relative to the instrument stiffness. We evaluated compliance effects of the ARES 2K-FRT on LVE measurements by systematically comparing the results of the frequency sweep mode obtained with 25 and 8 mm plates, respectively. In addition to the transducer, the test fixtures do contribute significantly to the system compliance. Without correction, the upper limit for the complex modulus vertical bar G*vertical bar is approximately close to 4 x 10(5) Pa at 10% uncertainty, when using 25 mm plates. This limit is lower than the plateau modulus (G(N)(0)) ofof most polymers. Therefore, instrument compliance can lead to significant errors for G(N)(0) and wrong scaling for G '' in the plateau and Rouse regions. The respective roles of transducer and tool compliances are discussed. The FRT transducer compliance is corrected in real time in the instrument firmware. Tool compliance is a common problem for all rheometers when measuring stiff samples. |
语种 | 英语 |
WOS记录号 | WOS:000296515600010 |
出版者 | SPRINGER |
源URL | [http://ir.iccas.ac.cn/handle/121111/73677] ![]() |
专题 | 中国科学院化学研究所 |
通讯作者 | Liu, Chen-Yang |
作者单位 | 1.Chinese Acad Sci, Beijing Natl Lab Mol Sci, CAS Key Lab Engn Plast, Joint Lab Polymer Sci & Mat,Inst Chem, Beijing 100190, Peoples R China 2.Catholic Univ Louvain, Unite Chim & Phys Hauts Polymeres, B-1348 Louvain, Belgium 3.TA Instruments, New Castle, DE 19720 USA 4.TA Instruments, D-65760 Eschborn, Germany |
推荐引用方式 GB/T 7714 | Liu, Chen-Yang,Yao, Minglong,Garritano, Ronald G.,et al. Instrument compliance effects revisited: linear viscoelastic measurements[J]. RHEOLOGICA ACTA,2011,50(5-6):537-546. |
APA | Liu, Chen-Yang,Yao, Minglong,Garritano, Ronald G.,Franck, Aloyse J.,&Bailly, Christian.(2011).Instrument compliance effects revisited: linear viscoelastic measurements.RHEOLOGICA ACTA,50(5-6),537-546. |
MLA | Liu, Chen-Yang,et al."Instrument compliance effects revisited: linear viscoelastic measurements".RHEOLOGICA ACTA 50.5-6(2011):537-546. |