Thermal analysis and expected performance of the low energy instrument on board the HXMT satellite
文献类型:期刊论文
作者 | Huo Jia![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
刊名 | CHINESE PHYSICS C
![]() |
出版日期 | 2010 |
卷号 | 34期号:12页码:1812-1817 |
关键词 | LE HXMT satellite attitude thermal calculation |
ISSN号 | 1674-1137 |
DOI | 10.1088/1674-1137/34/12/005 |
英文摘要 | The Low Energy X-ray Instrument (LE) of the Hard X-ray Modulation Telescope (HXMT) uses the Swept Charge Device (SCD) to detect the X-rays in 1-15 keV. The performance of SCD is vulnerable to temperature. We analyzed the thermal condition of LE at different satellite working attitudes with the Finite Element Method (FEM). It is shown that the angle between the sunlight and the normal line of the barrier should be less than 26 degrees, to keep the SCD detectors working in the required temperature range, i.e. -40 degrees C to -80 degrees C. We find that the performance of LE is very stable in this temperature range, with a typical energy resolution of 160 eV at 5.9 keV. |
WOS研究方向 | Physics, Nuclear ; Physics, Particles & Fields |
语种 | 英语 |
WOS记录号 | WOS:000285279100006 |
源URL | [http://ir.ihep.ac.cn/handle/311005/282223] ![]() |
专题 | 硬X射线调制望远镜卫星 |
推荐引用方式 GB/T 7714 | Huo Jia,Liu Xiao-Yan,Yang Yan-Ji,et al. Thermal analysis and expected performance of the low energy instrument on board the HXMT satellite[J]. CHINESE PHYSICS C,2010,34(12):1812-1817. |
APA | Huo Jia.,Liu Xiao-Yan.,Yang Yan-Ji.,Li Cheng-Kui.,Chen Tian-Xiang.,...&Zhang Jun-Min.(2010).Thermal analysis and expected performance of the low energy instrument on board the HXMT satellite.CHINESE PHYSICS C,34(12),1812-1817. |
MLA | Huo Jia,et al."Thermal analysis and expected performance of the low energy instrument on board the HXMT satellite".CHINESE PHYSICS C 34.12(2010):1812-1817. |
入库方式: OAI收割
来源:高能物理研究所
浏览0
下载0
收藏0
其他版本
除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。