MRI compatibility of several early transition metal based alloys and its influencing factors

文献类型:期刊论文

作者Zhou, DB; Wang, SG; Wang, SP; Ai, HJ; Xu, J; Ai, HJ (reprint author), China Med Univ, Sch Stomatol, Shenyang 110002, Liaoning, Peoples R China.; Xu, J (reprint author), Chinese Acad Sci, Shenyang Natl Lab Mat Sci, Inst Met Res, Shenyang 110016, Peoples R China.
刊名JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS
出版日期2018
卷号106期号:1页码:377-385
关键词Total Hip-arthroplasty Magnetic-susceptibility Artifact Reduction Vascular Stents Mechanical-properties Nb-60ta-2zr Alloy Dental Materials Zr-14nb Alloy Implants Strength
ISSN号1552-4973
英文摘要Magnetic resonance imaging (MRI) compatibility of three early transition metal (ETM) based alloys was assessed in vitro with agarose gel as a phantom, including Zr-20Nb, near-equiatomic (TiZrNbTa)(90)Mo-10 and Nb-60Ta-2Zr, together with pure tantalum and L605 Co-Cr alloy for comparison. The artifact extent in the MR image was quantitatively characterized according to the maximum area of 2D images and the total volume in reconstructed 3D images with a series of slices under acquisition by fast spin echo (FSE) sequence and gradient echo (GRE) sequence. It was indicated that the artifacts extent of L605 Co-Cr alloy with a higher magnetic susceptibility ((v)) was approximately 3-fold greater than that of the ETM-based alloys with (v) in the range of 160-250 ppm. In the ETM group, the MRI compatibility of the materials can be ranked in a sequence of Zr-20Nb, pure tantalum, (TiZrNbTa)(90)Mo-10 and Nb-60Ta-2Zr. In addition, using a rabbit cadaver with the implanted tube specimens as a model for ex vivo assessment, it was confirmed that the artifact severity of Nb-60Ta-2Zr alloy is significantly reduced in comparison with the L605 alloy. (c) 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 377-385, 2018.; Magnetic resonance imaging (MRI) compatibility of three early transition metal (ETM) based alloys was assessed in vitro with agarose gel as a phantom, including Zr-20Nb, near-equiatomic (TiZrNbTa)(90)Mo-10 and Nb-60Ta-2Zr, together with pure tantalum and L605 Co-Cr alloy for comparison. The artifact extent in the MR image was quantitatively characterized according to the maximum area of 2D images and the total volume in reconstructed 3D images with a series of slices under acquisition by fast spin echo (FSE) sequence and gradient echo (GRE) sequence. It was indicated that the artifacts extent of L605 Co-Cr alloy with a higher magnetic susceptibility ((v)) was approximately 3-fold greater than that of the ETM-based alloys with (v) in the range of 160-250 ppm. In the ETM group, the MRI compatibility of the materials can be ranked in a sequence of Zr-20Nb, pure tantalum, (TiZrNbTa)(90)Mo-10 and Nb-60Ta-2Zr. In addition, using a rabbit cadaver with the implanted tube specimens as a model for ex vivo assessment, it was confirmed that the artifact severity of Nb-60Ta-2Zr alloy is significantly reduced in comparison with the L605 alloy. (c) 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 377-385, 2018.
学科主题Engineering, Biomedical ; Materials Science, bioMaterials
语种英语
资助机构SYNL by the National Natural Science Foundation of China [51571192]
公开日期2018-06-05
源URL[http://ir.imr.ac.cn/handle/321006/79622]  
专题金属研究所_中国科学院金属研究所
通讯作者Ai, HJ (reprint author), China Med Univ, Sch Stomatol, Shenyang 110002, Liaoning, Peoples R China.; Xu, J (reprint author), Chinese Acad Sci, Shenyang Natl Lab Mat Sci, Inst Met Res, Shenyang 110016, Peoples R China.
推荐引用方式
GB/T 7714
Zhou, DB,Wang, SG,Wang, SP,et al. MRI compatibility of several early transition metal based alloys and its influencing factors[J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS,2018,106(1):377-385.
APA Zhou, DB.,Wang, SG.,Wang, SP.,Ai, HJ.,Xu, J.,...&Xu, J .(2018).MRI compatibility of several early transition metal based alloys and its influencing factors.JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS,106(1),377-385.
MLA Zhou, DB,et al."MRI compatibility of several early transition metal based alloys and its influencing factors".JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS 106.1(2018):377-385.

入库方式: OAI收割

来源:金属研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。