中国科学院机构知识库网格
Chinese Academy of Sciences Institutional Repositories Grid
A position-sensitive particle detector for Coulomb excitation experiment

文献类型:期刊论文

作者Toh, Y; Oshima, M; Hayakawa, T; Hatsukawa, Y; Katakura, J; Matsuda, M; Iimura, H; Kusakari, H; Nishimiya, D; Sugawara, M
刊名REVIEW OF SCIENTIFIC INSTRUMENTS
出版日期2002
卷号73页码:47-50
ISSN号0034-6748
英文摘要For Coulomb excitation experiments using a gamma -ray detector array, a position-sensitive particle detector system that has a good angular resolution, capability of managing a high-counting rate, robustness for a radiation damage, compactness (under 110 mm phi) and easiness of handling is necessary. For this purpose we developed a new device utilizing plastic and yttrium aluminum perovskite activated by cerium (formula YAlO3:Ce abbreviated YAP Ce) scintillators and a position-sensitive photomultiplier tube. The detector system could operate at a high-counting rate (about 10(5)), and with a good position resolution of 0.5-1.2 mm, which was enough for Doppler correction of the gamma rays. It could sustain the radiation damage caused by Coulomb excitation experiments with heavy ions for more than 50 days. The detector chamber was well fitted in the 110 mm phi diameter of the internal space of a compact gamma -ray detector array, GEMINI. This was much easier to handle than the position-sensitive gas counter. In Coulomb excitation experiment utilizing this device, the nucleus Ge-74 has been Coulomb excited using a Pb target. The transitional E2 matrix elements have been determined and the device has turned out to be useful in Coulomb excitation experiments with heavy-ion beams. (C) 2002 American Institute of Physics.
WOS关键词STATE
WOS研究方向Instruments & Instrumentation ; Physics
语种英语
WOS记录号WOS:000172906000008
出版者AMER INST PHYSICS
源URL[http://119.78.100.186/handle/113462/38285]  
专题中国科学院近代物理研究所
通讯作者Toh, Y
作者单位1.Japan Atom Energy Res Inst, Tokai, Ibaraki 3191195, Japan
2.Chiba Univ, Fac Educ, Chiba 2638522, Japan
3.Chiba Inst Technol, Chiba 2750023, Japan
4.Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
推荐引用方式
GB/T 7714
Toh, Y,Oshima, M,Hayakawa, T,et al. A position-sensitive particle detector for Coulomb excitation experiment[J]. REVIEW OF SCIENTIFIC INSTRUMENTS,2002,73:47-50.
APA Toh, Y.,Oshima, M.,Hayakawa, T.,Hatsukawa, Y.,Katakura, J.,...&Zhang, YH.(2002).A position-sensitive particle detector for Coulomb excitation experiment.REVIEW OF SCIENTIFIC INSTRUMENTS,73,47-50.
MLA Toh, Y,et al."A position-sensitive particle detector for Coulomb excitation experiment".REVIEW OF SCIENTIFIC INSTRUMENTS 73(2002):47-50.

入库方式: OAI收割

来源:近代物理研究所

浏览0
下载0
收藏0
其他版本

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。