The retarding Bessel-Box An electron-spectrometer designed for pump/probe experiments
文献类型:期刊论文
作者 | Schiwietz, G.1,2; Beye, M.1,2; Kuehn, D.1,2; Xiao, G.1,2,3 |
刊名 | JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA
![]() |
出版日期 | 2015-08-01 |
卷号 | 203页码:51-59 |
关键词 | Electron spectrometer Detection-solid-angle Energy resolution Femtosecond-pump/probe experiments Photo electrons Auger electrons |
ISSN号 | 0368-2048 |
DOI | 10.1016/j.elspec.2015.06.011 |
英文摘要 | A new type of electrostatic electron spectrometer is developed, capable of particular sensitive measurements of energy spectra and time-of-flight distributions. This instrument is specifically designed and optimized for laser-pump/X-ray-probe measurements, where photo electrons or Auger electrons from surfaces, clusters, molecular or atomic targets are being measured with high time-resolution at an extremely low detection-noise level. The compact and robust cylinder-symmetrical system is a strongly improved Bessel-Box design, featuring electron retardation, a large detection solid-angle, about 100% electron transmission (gridless design) and excellent time-resolution. In this paper we describe the principle of operation of this type of spectrometer and various tests. We present quantitative results for electron measurements with different solid-state targets and two different electron-detection systems in comparison to electron-trajectory simulations inside the electrostatic spectrometer fields. Picosecond-pump/probe operation has been tested with high laser power and even the ability to work under femtosecond-pump-probe conditions with electron detection at the BESSY II slicing facility has been proven. (C) 2015 Elsevier B.V. All rights reserved. |
WOS关键词 | ENERGY ANALYZER ; NEW-GENERATION ; LASER |
WOS研究方向 | Spectroscopy |
语种 | 英语 |
WOS记录号 | WOS:000360512000009 |
出版者 | ELSEVIER SCIENCE BV |
源URL | [http://119.78.100.186/handle/113462/40523] ![]() |
专题 | 中国科学院近代物理研究所 |
通讯作者 | Schiwietz, G. |
作者单位 | 1.Helmholtz Zentrum Berlin Mat & Energie GmbH, Inst Methoden & Instrumentierung Forsch Synchrotr, D-14109 Berlin, Germany 2.Helmholtz Zentrum Berlin Mat & Energie GmbH, Div HMI SF8, D-14109 Berlin, Germany 3.Inst Modern Phys, Lanzhou 730000, Peoples R China |
推荐引用方式 GB/T 7714 | Schiwietz, G.,Beye, M.,Kuehn, D.,et al. The retarding Bessel-Box An electron-spectrometer designed for pump/probe experiments[J]. JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA,2015,203:51-59. |