An accurate low current measurement circuit for heavy iron beam current monitor

文献类型:期刊论文

作者Zhou, Chao-Yang1,2; Su, Hong1; Mao, Rui-Shi1; Dong, Cheng-Fu1; Qian, Yi1; Kong, Jie1
刊名NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS
出版日期2012-06-01
卷号280页码:84-87
关键词Faraday Cup Low Beam Current Monitor Picoammeter Gated Integrator
ISSN号0168-583X
DOI10.1016/j.nimb.2012.01.033
文献子类Article
英文摘要Heavy-ion beams at 10(6) particles per second have been applied to the treatment of deep-seated inoperable tumors in the therapy terminal of the Heavy Ion Research Facility in Lanzhou (HIRFL) which is located at the Institute of Modern Physics, Chinese Academy of Sciences (IMP, CAS). An accurate low current measurement circuit following a Faraday cup was developed to monitor the beam current at pA range. The circuit consisted of a picoammeter with a bandwidth of 1 kHz and a gated integrator (GI). A low input bias current precision amplifier and new guarding and shielding techniques were used in the picoammeter circuit which allowed as to measure current less than 1 pA with a current gain of 0.22 V/pA and noise less than 10 fA. This paper will also describe a novel compensation approach which reduced the charge injection from switches in the GI to 10(-18) C, and a T-switch configuration which was used to eliminate leakage current in the reset switch. (C) 2012 Elsevier B.V. All rights reserved.
资助项目CAS[KJCX2-YW-N27]
WOS研究方向Instruments & Instrumentation ; Nuclear Science & Technology ; Physics
语种英语
WOS记录号WOS:000304336200012
出版者ELSEVIER SCIENCE BV
源URL[http://119.78.100.186/handle/113462/48889]  
专题中国科学院近代物理研究所
通讯作者Su, Hong
作者单位1.Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
2.Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
推荐引用方式
GB/T 7714
Zhou, Chao-Yang,Su, Hong,Mao, Rui-Shi,et al. An accurate low current measurement circuit for heavy iron beam current monitor[J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS,2012,280:84-87.
APA Zhou, Chao-Yang,Su, Hong,Mao, Rui-Shi,Dong, Cheng-Fu,Qian, Yi,&Kong, Jie.(2012).An accurate low current measurement circuit for heavy iron beam current monitor.NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS,280,84-87.
MLA Zhou, Chao-Yang,et al."An accurate low current measurement circuit for heavy iron beam current monitor".NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS 280(2012):84-87.