Study of detector unit of Micro-PET
文献类型:期刊论文
作者 | Ding, Xianli; Wang, Jinchuan![]() ![]() ![]() ![]() ![]() ![]() |
刊名 | He Jishu/Nuclear Techniques
![]() |
出版日期 | 2006 |
卷号 | 29页码:388-391 |
ISSN号 | 0253-3219 |
英文摘要 | This paper presents a series of tests of a detector unit which will be used in a fast, accurate, convenient and inexpensive medical diagnostic instrument for tumor-Micro Positron Emission Tomography (Micro-PET). The structure of the detector unit, its network of readout, the electronics and the data acquisition system are described. The position resolution reaches 1.02 mm. The ratio of the anode output maximum to minimum given by the 64 channels of H8500 is 3.1, demonstrating a rather good uniformity. |
出版者 | Science Press, Beijing, China |
源URL | [http://119.78.100.186/handle/113462/64132] ![]() |
专题 | 中国科学院近代物理研究所 |
作者单位 | 1.Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China 2.Graduate School, Chinese Academy of Sciences, Beijing 100049, China |
推荐引用方式 GB/T 7714 | Ding, Xianli,Wang, Jinchuan,Qi, Huirong,et al. Study of detector unit of Micro-PET[J]. He Jishu/Nuclear Techniques,2006,29:388-391. |
APA | Ding, Xianli.,Wang, Jinchuan.,Qi, Huirong.,Xiao, Guoqing.,Xiao, Zhigang.,...&Guo, Zhongyan.(2006).Study of detector unit of Micro-PET.He Jishu/Nuclear Techniques,29,388-391. |
MLA | Ding, Xianli,et al."Study of detector unit of Micro-PET".He Jishu/Nuclear Techniques 29(2006):388-391. |
入库方式: OAI收割
来源:近代物理研究所