英文摘要This paper describes the current status, working principles, design requirement and method for high density multi-channel front-end electronics and multi-channel charge sensitive preamplifier of array detector briefly. Specially a new 48-channel charge sensitive preamplifier developed for the silicon strip array detector system in IMP is introduced, which has low cost, high density, 1% energy resolution and good stability.