ACTIVATION OF A 250 MEV SC-CYCLOTRON FOR PROTON THERAPY

文献类型:研究报告

作者G. Steen ; B. Amrein ; N. Frey ; P. Frey ; D. Kiselev ; M. Kostezer ; R. Lüscher ; D. Mohr ; O. Morath ; A. Schmidt ; M. Wohlmuther ; J.M. Schippers
完成日期2010-09-10
关键词Psi’s 250 Mev Sc-cyclotron Proton Therapy Mcnpx
英文摘要Beam losses in dedicated 230-250 MeV cyclotrons for
proton therapy cause radioactivity in machine parts. A
systematic study has been performed of the activation of
PSI’s 250 MeV SC-cyclotron for proton therapy. Since
the start of the cyclotron operation dose rate
measurements have been made as a function of time at
several locations in and around the cyclotron. Gamma ray
spectra have been measured of selected iron inserts in the
pole and of copper disks in the liner of the RF system.
The isotopic composition of the activation has been
derived and compared with activations calculated with
Monte Carlo calculations (MCNPX). The data and beam
history of the cyclotron allow predictions of the dose rate
during service activities shortly after beam interruption as
well as after a specified period of operation.