FAST SCANNING TECHNIQUES FOR CANCER THERAPY WITH HADRONS – A DOMAIN OF CYCLOTRONS
文献类型:会议论文
作者 | J.M. Schippers ; D. Meer ; E. Pedroni |
出版日期 | 2010-09-10 |
会议日期 | 2010-09-06 |
会议地点 | 兰州 |
英文摘要 | In protontherapy fast 3D pencil beam scanning is regarded as the most optimal dose delivery method. The requirements to apply this treatment technique and to obtain the maximum possible benefit have a big impact on the accelerator concept. Routinely a very stable, reproducible and adjustable beam intensity is needed, which can be set at a few percent accuracy within a millisecond. Quick changes of maximum intensity from the cyclotron are also needed when changing treatment room. Rescanning the tumour volume at high speed to prevent motion artefacts, needs beam energy variations within 50-80 ms. It will be shown that a cyclotron offers the most advantageous possibilities to achieve this ambitious performance |